Cardiorespiratory Transitioning at Birth during Varying Levels of Resuscitation and the Role of Dry-Electrode ECG (NeoBeat) and Pulse Oximetry in a Low Resource Setting
Neonatology · 13.05.2026 · 1-10
DOI: 10.1159/000552554 ↗Kurz gefasst
Fachartikel in der Zeitschrift Neonatology, online erschienen am 13.05.2026. Verfasst von Emily Ahn und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Neonatology, NewYork-Presbyterian-Weill Cornell Medicine, New York, New York, USA, ema9066@med.cornell.edu. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NCATS NIH HHS.
Themen: Reanimation
Abstract (Originalsprache)
Introduction
Studies on neonatal cardiorespiratory transition are important in guiding resuscitation guidelines. The objectives were to determine the immediate transitional changes of heart rate (HR) and saturation values in neonates who received various levels of resuscitation and the level of agreement between NeoBeat and pulse oximetry.
Methods
High-risk deliveries (n = 63) in a low resource setting were observed using pulse oximetry and NeoBeat in three groups: comfortably breathing receiving no resuscitation, labored breathing receiving oxygen only, and apneic receiving bag mask ventilation (BMV) plus oxygen. Data were analyzed by time of birth and compared for levels of agreement.
Results
Comfortably breathing neonates had initial saturation values (86 ± 13%) with relatively high initial HRs (177 ± 19 beats per minute [bpm]). For neonates receiving oxygen only, 43% reached target saturation goals despite an initial HR of 150 ± 21 bpm, while the BMV group had 73% reach target saturation goals with an initial HR of 88 ± 40 bpm. Bland-Altman revealed low bias in all groups and progressively wider lines of agreement as neonates required more resuscitation. In the BMV group, pulse oximetry often displayed a lower HR than NeoBeat at lower mean HR values.
Conclusions
Elevated early saturation values and higher HRs in comfortably breathing neonates may be indicative of a seamless cardiorespiratory adaptation. For neonates requiring resuscitation, oxygen alone without positive end-expiratory pressure may delay respiratory transitioning, as evidenced by low oxygen saturations at 5 min, whereas a higher percentage of neonates receiving BMV with oxygen met target saturation goals.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Neonatology, NewYork-Presbyterian-Weill Cornell Medicine, New York, New York, USA, ema9066@med.cornell.edu.
- Department of Obstetrics and Gynecology, Kilimanjaro Christian Medical Centre, Moshi, Tanzania.
- Department of Pediatrics, Kilimanjaro Christian Medical Centre, Moshi, Tanzania.
- KCMC University, Moshi, Tanzania.
- Division of Neonatology, NewYork-Presbyterian-Weill Cornell Medicine, New York, New York, USA.
Förderung
- NCATS NIH HHS
Zitieren
APA
Ahn, E., Cypriane, J., Ngowi, E., Shayo, A., & Perlman, J. (2026). Cardiorespiratory Transitioning at Birth during Varying Levels of Resuscitation and the Role of Dry-Electrode ECG (NeoBeat) and Pulse Oximetry in a Low Resource Setting. Neonatology, 1-10. https://doi.org/10.1159/000552554
Vancouver
Ahn E, Cypriane J, Ngowi E, Shayo A, Perlman J. Cardiorespiratory Transitioning at Birth during Varying Levels of Resuscitation and the Role of Dry-Electrode ECG (NeoBeat) and Pulse Oximetry in a Low Resource Setting. Neonatology. 2026:1-10. doi:10.1159/000552554
BibTeX
@article{Ahn2026,
author = {Ahn, Emily and Cypriane, Jackline and Ngowi, Ester and Shayo, Aisa and Perlman, Jeffrey},
title = {Cardiorespiratory Transitioning at Birth during Varying Levels of Resuscitation and the Role of Dry-Electrode ECG (NeoBeat) and Pulse Oximetry in a Low Resource Setting},
journal = {Neonatology},
year = {2026},
pages = {1--10},
doi = {10.1159/000552554}
}
EUVIMED