Fachartikel in der Zeitschrift Stereotactic and Functional Neurosurgery, online erschienen am 28.04.2026. Verfasst von Michael G Hart und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Psychology and Neuroscience, St George's School of Health and Medical Sciences, City St George's, University of London, London, UK, mic…. Die Arbeit nennt 14 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Targeting and coordinate derivation are critical steps in deep brain stimulation (DBS). Multiple planning stations are available to assist in this process, yet it is unclear how they compare in their methods and outputs. We hypothesised that planning DBS would generate the same coordinates independent of the planning station used.

Methods

We compared the planning of DBS to the subthalamic nucleus (STN) using three planning stations (Brainlab Elements v2, Medtronic StealthStation™ S8, and Renishaw neuroinspire™ v6.2.2). Six users (2 consultants/attendings, 2 senior fellows, 2 clinical lecturers/senior residents) planned bilateral STN DBS for the same 10 randomly ordered subjects on each planning station in 3 spaces (native T2 image, AC-PC, stereotactic frame), generating 3,240 data points. Differences in individual XYZ coordinates were analysed per space using parametric statistical tests.

Results

Mean coordinates in image space varied by 0.33 mm (Z-plane) between planning stations, with statistically significant differences in the X-plane (Brainlab) and Y- and Z-planes (Renishaw). Mean coordinates in AC-PC space varied by a maximum of 1.02 mm (Z-plane) with significant differences present in all 3 planes for Renishaw. Mean coordinates in stereotactic frame space varied by 0.53 mm (Z-plane) with differences present in the X-plane (Brainlab), Y-plane (Renishaw), and Z-plane (all). Between-rater and between-subject analyses did not reveal any significant differences.

Conclusion

Differences in derived coordinates between planning station are present but small (mean

Schlagwörter

Stichwörter der Autor:innen

Deep brain stimulationPlanning station softwareStereotaxic techniquesSubthalamic nucleusTarget coordinates

Autor:innen und Institutionen

  • Department of Psychology and Neuroscience, St George's School of Health and Medical Sciences, City St George's, University of London, London, UK, michael.hart8@nhs.net.
  • Department of Psychology and Neuroscience, St George's School of Health and Medical Sciences, City St George's, University of London, London, UK, rkhataybeh@gmail.com.
  • Department of Psychology and Neuroscience, St George's School of Health and Medical Sciences, City St George's, University of London, London, UK.

Zitieren

APA

Hart, M. G., Alkhataybeh, R. A., Shah, R. S., Shaikh, A., Mostofi, A., & Pereira, E. A. C. (2026). Choice of Neurosurgical Planning Software Affects Deep Brain Stimulation Target Coordinate Generation. Stereotactic and Functional Neurosurgery, 1-9. https://doi.org/10.1159/000551905

Vancouver

Hart MG, Alkhataybeh RA, Shah RS, Shaikh A, Mostofi A, Pereira EAC. Choice of Neurosurgical Planning Software Affects Deep Brain Stimulation Target Coordinate Generation. Stereotactic and Functional Neurosurgery. 2026:1-9. doi:10.1159/000551905

BibTeX
@article{Hart2026,
  author = {Hart, Michael G and Alkhataybeh, Redab Ahmad and Shah, Rahul S and Shaikh, Adnan and Mostofi, Abteen and Pereira, Erlick A C},
  title = {Choice of Neurosurgical Planning Software Affects Deep Brain Stimulation Target Coordinate Generation},
  journal = {Stereotactic and Functional Neurosurgery},
  year = {2026},
  pages = {1--9},
  doi = {10.1159/000551905}
}
Kostenlos starten →