Prognostic Value of Remote Monitoring Alarms for Technique Failure in Automated Peritoneal Dialysis
Blood Purification · 02.04.2026 · 1-9
DOI: 10.1159/000551828 ↗Kurz gefasst
Fachartikel in der Zeitschrift Blood Purification, online erschienen am 02.04.2026. Verfasst von Kuan-Chiao Huang und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nephrology, Department of Internal Medicine, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University…. Hauptthemen laut MEDLINE: Peritoneal Dialysis, Clinical Alarms. Die Arbeit nennt 25 Literaturangaben.
Abstract (Originalsprache)
Introduction
While remote patient monitoring (RPM) is associated with decreased hospitalization and technique failure in automated peritoneal dialysis (APD), the prognostic utility of individual alarm patterns remains unclear.
Methods
We performed a retrospective, single-center cohort study of adult APD patients managed with the Sharesource RPM platform at National Cheng Kung University Hospital. Three predefined red-flag alarms, including initial drain variance >50%, lost therapy volume >10%, and lost treatment time >30 min, were recorded from September 2019 to May 2020. Alarm burden was expressed as episodes per patient-month and categorized into tertiles. Patients were followed for 3 years for death-censored technique failure, defined as permanent transfer to hemodialysis. Associations were evaluated using logistic regression. Fine and Gray competing-risk models and generalized estimating equations served as sensitivity analyses.
Results
Sixty-seven patients (mean age 45 ± 13 years; 61.2% male) generated 272 alarm episodes. The initial drain variance accounted for 54.8% of alarms, lost treatment time for 39.3%, and lost therapy volume for 5.9%. Technique failure occurred in 8 patients (11.9%), with rates rising across alarm-frequency tertiles (4.5%, 9.1%, and 21.7%; p for trend = 0.074). In multivariable analysis, initial drain variance > 50% independently predicted technique failure (adjusted odds ratio [OR] 1.48; 95% confidence interval [CI]: 1.13–2.04; p = 0.006). Results were consistent in sensitivity analyses (Fine and Gray subdistribution hazard ratio 1.55, 95% CI: 1.07–2.25; GEE OR 2.58, 95% CI: 1.00–6.62). The other two alarm types were not significantly associated with outcomes.
Conclusion
An RPM alarm for initial drain variance >50% emerged as an early marker of technique failure in APD. Incorporating alarm-pattern analysis into routine RPM review may facilitate targeted surveillance and timely intervention for high-risk patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Nephrology, Department of Internal Medicine, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan.
- Institute of Clinical Medicine, College of Medicine, National Cheng Kung University, Tainan, Taiwan.
- Institute of Clinical Pharmacy and Pharmaceutical Sciences, College of Medicine, National Cheng Kung University, Tainan, Taiwan.
- Division of Nephrology, Department of Internal Medicine, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan, heroccl1113@gmail.com.
- Institute of Clinical Medicine, College of Medicine, National Cheng Kung University, Tainan, Taiwan, heroccl1113@gmail.com.
Zitieren
APA
Huang, K. C., Chao, J. Y., Liu, K. H., Lin, W. R., Sung, J. M., Wang, M. C., Tseng, C. C., & Lee, C. C. (2026). Prognostic Value of Remote Monitoring Alarms for Technique Failure in Automated Peritoneal Dialysis. Blood Purification, 1-9. https://doi.org/10.1159/000551828
Vancouver
Huang KC, Chao JY, Liu KH, Lin WR, Sung JM, Wang MC, et al. Prognostic Value of Remote Monitoring Alarms for Technique Failure in Automated Peritoneal Dialysis. Blood Purification. 2026:1-9. doi:10.1159/000551828
BibTeX
@article{Huang2026,
author = {Huang, Kuan-Chiao and Chao, Jo-Yen and Liu, Kuan-Hung and Lin, Wei-Ren and Sung, Junne-Ming and Wang, Ming-Cheng and Tseng, Chin-Chung and Lee, Chia-Chun},
title = {Prognostic Value of Remote Monitoring Alarms for Technique Failure in Automated Peritoneal Dialysis},
journal = {Blood Purification},
year = {2026},
pages = {1--9},
doi = {10.1159/000551828}
}
EUVIMED