Assessing the Significance of Abnormal Inserted Perimodiolar Slim Electrodes: Perception and Management Strategies
Audiology and Neurotology · 2025
DOI: 10.1159/000546928 ↗Kurz gefasst
Fachartikel in der Zeitschrift Audiology and Neurotology, 2025. Verfasst von Phoebe Helena Ramos und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology of the University of Melbourne, Melbourne, Victoria, Australia, phoeberamos@gmail.com. Hauptthemen laut MEDLINE: Cochlear Implantation, Cochlear Implants, Speech Perception, Electrodes, Implanted. Die Arbeit nennt 26 Literaturangaben.
Abstract (Originalsprache)
Introduction
Slim perimodiolar electrodes (PSEs) are designed for close modiolar proximity, but deviations from ideal placement may affect speech perception. This study investigates the prevalence, radiologic characteristics, and audiologic outcomes of abnormally positioned PSE arrays and proposes intraoperative management strategies.
Methods
We retrospectively reviewed 239 adults implanted with Cochlear® PSE arrays (2015–2023). Postoperative CT scan identified arrays positioned laterally to mid-scala without tip fold-over. Objective metrics such as angular insertion depth (AID), intracochlear position index (ICPI), wrapping factor (WF), and maximum insertion angle were compared to a control group with normal positioning (n = 52). Speech perception outcomes at 3 and 12 months were assessed using monosyllabic word tests and compared to a reference cohort (n = 1,414).
Results
Abnormal positioning occurred in 3% (n = 7). All arrays remained within scala tympani but had shallower insertions (AID 296° vs. 392°), greater lateralization (ICPI 0.52 vs. 0.33), and higher WF (0.62 vs. 0.36). Compared to controls, speech scores were lower at 3 months (33% vs. 65%) and 12 months (45% vs. 74%). Intraoperative X-rays did not detect these deviations.
Conclusion
Abnormal PSE placement is uncommon but associated with poorer speech outcomes. Surgeons should assess electrode shape intraoperatively and consider reinsertion if the classic curl is absent. Postoperative CT remains critical for quality assurance. Tailored intraoperative strategies are needed to prevent lateralization and optimize outcomes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology of the University of Melbourne, Melbourne, Victoria, Australia, phoeberamos@gmail.com.
- Department of Otolaryngology of the Pontificia Universidad Católica de Chile, Santiago, Chile, phoeberamos@gmail.com.
- Royal Victorian Eye and Ear Hospital, Melbourne, Victoria, Australia, phoeberamos@gmail.com.
- Department of Otolaryngology of the University of Melbourne, Melbourne, Victoria, Australia.
- Royal Victorian Eye and Ear Hospital, Melbourne, Victoria, Australia.
Zitieren
APA
Ramos, P. H., Collins, A., Gerard, J. M., & Gersdorff, G. (2025). Assessing the Significance of Abnormal Inserted Perimodiolar Slim Electrodes: Perception and Management Strategies. Audiology and Neurotology. https://doi.org/10.1159/000546928
Vancouver
Ramos PH, Collins A, Gerard JM, Gersdorff G. Assessing the Significance of Abnormal Inserted Perimodiolar Slim Electrodes: Perception and Management Strategies. Audiology and Neurotology. 2025. doi:10.1159/000546928
BibTeX
@article{Ramos2025,
author = {Ramos, Phoebe Helena and Collins, Aaron and Gerard, Jean-Marc and Gersdorff, Guillaume},
title = {Assessing the Significance of Abnormal Inserted Perimodiolar Slim Electrodes: Perception and Management Strategies},
journal = {Audiology and Neurotology},
year = {2025},
doi = {10.1159/000546928}
}