Application of Remote Dielectric Sensing in Volumetric Evaluation for Maintenance Hemodialysis Patients
Blood Purification · 2025
DOI: 10.1159/000546872 ↗Kurz gefasst
Fachartikel in der Zeitschrift Blood Purification, 2025. Verfasst von Yuqing Chen und 2 weiteren Autor:innen. Beteiligte Einrichtung: Renal Division, Department of Medicine, Peking University First Hospital, Beijing, China. Hauptthemen laut MEDLINE: Renal Dialysis, Remote Sensing Technology. Die Arbeit nennt 27 Literaturangaben.
Abstract (Originalsprache)
Introduction
Fluid overload increases the risk of cardiovascular events and all-cause mortality in hemodialysis patients. Remote dielectric sensing (ReDS) is a newly wearable device which can evaluate the volumetric status via the ratio of fluid and air in pulmonary tissue. In this study we explored whether ReDS can be used in the evaluation of volume status in dialysis patients.
Methods
We involved maintenance hemodialysis patients in the cross-sectional study and obtained both pre- and post-dialysis ReDS values. We also collected pre- and post-dialysis overhydration values by bioelectrical impedance analysis and biochemistry indicators like brain natriuretic peptide and analyzed their correlations with ReDS values to explore whether ReDS was feasible in fluid assessment for hemodialysis patients.
Results
115 maintenance hemodialysis patients were enrolled. The average pre-dialysis ReDS reading was 27.47% ± 4.88% (range from 16.00% to 39.00%), and the median post-dialysis ReDS value was 24% (range from 16% to 37%). ReDS can distinguish fluid overload patients (pre-dialysis overhydration value ≥2.5 L, p < 0.001). ReDS value has a positive correlation with overhydration value (pre-dialysis: R = 0.38, p < 0.001) and a negative correlation with left ventricular ejection fraction (R = −0.333, p = 0.018). ROC analysis showed that cut-off of pre-dialysis ReDS value to discriminate pre-dialysis overhydration value ≥2.5 L was 26.5%.
Conclusion
ReDS has potential in fluid evaluation for maintenance hemodialysis patients and can serve as a useful tool to assist in clinical hydration status assessment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Renal Division, Department of Medicine, Peking University First Hospital, Beijing, China.
- Institute of Nephrology, Peking University, Beijing, China.
- Key Laboratory of Renal Disease, Ministry of Health of China, Beijing, China.
- Key Laboratory of Chronic Kidney Disease Prevention and Treatment, Ministry of Education, Beijing, China.
- Research Units of Diagnosis and Treatment of Immune-mediated Kidney Diseases, Chinese Academy of Medical Sciences, Beijing, China.
- Renal Division, Department of Medicine, Peking University First Hospital, Beijing, China, adventofmay@outlook.com.
Zitieren
APA
Chen, Y., Wang, C. Y., & Zhou, L. (2025). Application of Remote Dielectric Sensing in Volumetric Evaluation for Maintenance Hemodialysis Patients. Blood Purification. https://doi.org/10.1159/000546872
Vancouver
Chen Y, Wang CY, Zhou L. Application of Remote Dielectric Sensing in Volumetric Evaluation for Maintenance Hemodialysis Patients. Blood Purification. 2025. doi:10.1159/000546872
BibTeX
@article{Chen2025,
author = {Chen, Yuqing and Wang, Chun-Yue and Zhou, Li},
title = {Application of Remote Dielectric Sensing in Volumetric Evaluation for Maintenance Hemodialysis Patients},
journal = {Blood Purification},
year = {2025},
doi = {10.1159/000546872}
}
EUVIMED