A New Anti-Parietal Cell Antibody Titer Measurement Kit for Diagnosis of Autoimmune Gastritis
Digestion · 2025
DOI: 10.1159/000545454 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Digestion, 2025. Verfasst von Masanori Ito und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of General Internal Medicine, Hiroshima University Hospital, Hiroshima, Japan. Hauptthemen laut MEDLINE: Gastritis, Autoimmune Diseases, Parietal Cells, Gastric. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Introduction
Recent findings revealed that autoimmune gastritis (AIG) is not rare in Japan. Therefore, the accurate diagnosis of AIG is essential in upper gastrointestinal practice. Diagnostic criteria for AIG were established in 2023; however, the conventional fluorescent antibody (FA) method for anti-parietal cell antibody (APCA) titer measurements has low accuracy and is clinically problematic.
Methods
We developed a latex agglutination (LA) method using a new human-derived antigen. Samples subjected to measurements were sera from AIG cases (127 cases, 49 males, average age 65.9 years) and control cases (129 cases, 81 males, average age 66.1 years) provided by the main facility and three affiliated facilities in Japan. APCA in whole serum was measured using the FA and LA methods.
Results
The diagnostic ability of AIG using the FA method was as follows: sensitivity of 99.2%, specificity of 41.1%, and an overall agreement rate of 69.9%, with low specificity previously being reported. On the other hand, the new LA method had good diagnostic performance with a sensitivity of 93.7%, a specificity of 95.4%, and an overall agreement rate of 94.5%.
Conclusion
We developed a new APCA measurement system that will contribute to the accurate diagnosis of AIG. The use of this measurement system in clinical practice will facilitate the diagnosis and treatment of AIG.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of General Internal Medicine, Hiroshima University Hospital, Hiroshima, Japan.
- Department of Gastroenterology, Fujieda Municipal General Hospital, Shizuoka, Japan.
- Department of Internal Medicine, Kakogawa Central City Hospital, Kakogawa, Japan.
- Medical System Development Center, FUJIFILM Corporation, Amagasaki, Japan.
- Diagnostics Development Department, FUJIFILM Wako Pure Chemical Corporation, Osaka, Japan.
- Department of General Internal Medicine 2, Kawasaki Medical School General Medical Center, Okayama, Japan.
Zitieren
APA
Ito, M., Ito, M., Maruyama, Y., Terao, S., Okubo, K., Hidaka, K., Sakamoto, T., & Haruma, K. (2025). A New Anti-Parietal Cell Antibody Titer Measurement Kit for Diagnosis of Autoimmune Gastritis. Digestion. https://doi.org/10.1159/000545454
Vancouver
Ito M, Ito M, Maruyama Y, Terao S, Okubo K, Hidaka K, et al. A New Anti-Parietal Cell Antibody Titer Measurement Kit for Diagnosis of Autoimmune Gastritis. Digestion. 2025. doi:10.1159/000545454