A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study
Neonatology · 2025
DOI: 10.1159/000543413 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Neonatology, 2025. Verfasst von Francesco Cresi und 9 weiteren Autor:innen. Beteiligte Einrichtung: Neonatal Intensive Care Unit, Department of Public Health and Pediatrics, University of Turin, Turin, Italy. Hauptthemen laut MEDLINE: Infant, Premature, Noninvasive Ventilation, Intermittent Positive-Pressure Ventilation, Respiratory Distress Syndrome, Newborn. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 14 Literaturangaben.
Themen: Neonatologie
Abstract (Originalsprache)
Introduction
Nasal synchronized intermittent positive pressure ventilation (nSIPPV) is an effective non-invasive ventilation technique, especially for preterm infants. Patient-ventilator synchrony is essential for providing effective respiratory support; however, no automated system is currently available for monitoring this parameter. A new tool for automatic assessment of patient-ventilator synchrony, the SyncNIV system, was developed and applied in this pilot study to evaluate differences between nSIPPV and non-synchronized nasal intermittent positive pressure ventilation (nIPPV) in preterm infants with respiratory distress.
Methods
This study involved designing a custom algorithm for signal analysis. Data were collected through a polygraph that could simultaneously gather respiratory data from the patients and the ventilator. Patient-ventilator synchrony was evaluated by applying the SyncNIV system in a randomized crossover study designed to compare nSIPPV and nIPPV. The primary outcome was the mean instant Synchrony Index (i-SI), defined as the portion of the inspiration effort sustained by ventilator inflation, expressed as a percentage.
Results
Fourteen infants with a median (IQR) gestational age of 28.6 (25.6–30.3) were enrolled. We analyzed 43,304 ventilator inflations and 50,221 patient breaths. The i-SI was 54.69% (44.49–60.09) in nSIPPV and 39.54% (33.40–48.75) in nIPPV, p < 0.05.
Conclusion
The SyncNIV system confirmed better i-SI during nSIPPV than during nIPPV, demonstrating its effectiveness in assessing the differences between these two modes of non-invasive ventilation in preterm infants. The SyncNIV system could be a useful tool for optimizing the ventilation parameters and improving the effectiveness and comfort of respiratory support systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Neonatal Intensive Care Unit, Department of Public Health and Pediatrics, University of Turin, Turin, Italy.
- Città della Salute e della Scienza di Torino, Turin, Italy.
- Department of Maternal and Child Health, Policlinico Umberto I, Sapienza University of Rome, Rome, Italy.
Zitieren
APA
Cresi, F., Maggiora, E., Rubino, C., Ferroglio, M., Ruzzante, E., Piga, E., Giraudo, I., Limone, M., Terrin, G., & Coscia, A. (2025). A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study. Neonatology. https://doi.org/10.1159/000543413
Vancouver
Cresi F, Maggiora E, Rubino C, Ferroglio M, Ruzzante E, Piga E, et al. A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study. Neonatology. 2025. doi:10.1159/000543413
BibTeX
@article{Cresi2025,
author = {Cresi, Francesco and Maggiora, Elena and Rubino, Carlotta and Ferroglio, Mattia and Ruzzante, Elena and Piga, Enrico and Giraudo, Isaac and Limone, Marco and Terrin, Gianluca and Coscia, Alessandra},
title = {A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study},
journal = {Neonatology},
year = {2025},
doi = {10.1159/000543413}
}
EUVIMED