EUVIMEDEUVIMED entdecken
VergleichsstudieKohortenstudieOpen Access · CC BY-NC 4.0

Fine-Needle Aspiration versus the CytoCore ® Motorized Rotating Needle Device for Thyroid Nodule Biopsies: A Retrospective Cohort Study

Adarsh Verma, Rhonda McDowell, Anthony Porreca

Acta Cytologica · 2024

DOI: 10.1159/000541374 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Acta Cytologica, 2024. Verfasst von Adarsh Verma und 2 weiteren Autor:innen. Beteiligte Einrichtung: Division of Interventional Radiology, Department of Radiology, Morton Plant Mease Hospital, BayCare Health System, Clearwater, Florida, USA. Hauptthemen laut MEDLINE: Thyroid Nodule. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Recently, an FDA cleared motorized fine-needle aspiration device (CytoCore®, Praxis Medical) has become available which is designed to reduce sample variability by enabling more consistent sampling due to the rotational drilling action of the device in combination with the standard in and out motion used to access the thyroid nodule with a needle. The rotation of the needle permits the ability to collect a higher quantity of intact cellular material, which is optimal for determining adequacy and, ultimately, for making a diagnosis. The present study compares the diagnostic performance of a motorized fine needle aspiration (FNA) device to a historical cohort of patients biopsied using ultrasound-guided fine needle aspiration (US-FNA).

Methods

Data from 120 patients with thyroid nodules biopsied using a motorized FNA device was retrospectively analyzed. Patient demographics, lesion characteristics, number of passes, Bethesda category, and cellularity scores were compared to a historical control cohort of 100 patients who underwent US-FNA. Nondiagnostic and indeterminate samples rates for motorized FNA were separately compared to literature controls.

Results

A significantly reduced median number of passes were required with motorized FNA compared to US-FNA (1.48 ± 0.62 vs. 2.64 ± 1.63, p < 0.001). Adequate samples were obtained after the first pass for 58% of biopsies with motorized FNA compared to 11% with US-FNA. The cumulative percentage of adequate samples increased to 98% after two passes for motorized FNA versus 58% for the US-FNA group. The mean cellularity score was also significantly greater for motorized FNA (3.42 ± 0.63 vs. 1.9 ± 0.59; p < 0.001). A determinant diagnosis was possible for a greater number of samples in the motorized FNA group compared to the control group (91.6% vs. 78%; p = 0.05). The motorized FNA also had a lower nondiagnostic rate compared to US-FNA (2.0% vs. 10%) and a lower indeterminate rate compared to published rates associated with the use of FNA (8.3% vs. 20.0%; p = 0.05).

Conclusion

The motorized FNA device required less passes to obtain an adequate biopsy than US-FNA. Its use is also associated with obtaining samples with a higher cellularity and lower nondiagnostic and indeterminate sample rates.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Thyroid NoduleHumansRetrospective StudiesFemaleMaleMiddle AgedBiopsy, Fine-NeedleAgedAdultThyroid GlandEquipment DesignPredictive Value of TestsUltrasonography, InterventionalAged, 80 and overNeedles

Stichwörter der Autor:innen

BiopsyFine-needle aspirationIndeterminateNondiagnosticThyroid

Autor:innen und Institutionen

  • Division of Interventional Radiology, Department of Radiology, Morton Plant Mease Hospital, BayCare Health System, Clearwater, Florida, USA.

Zitieren

APA

Verma, A., McDowell, R., & Porreca, A. (2024). Fine-Needle Aspiration versus the CytoCore ® Motorized Rotating Needle Device for Thyroid Nodule Biopsies: A Retrospective Cohort Study. Acta Cytologica. https://doi.org/10.1159/000541374

Vancouver

Verma A, McDowell R, Porreca A. Fine-Needle Aspiration versus the CytoCore ® Motorized Rotating Needle Device for Thyroid Nodule Biopsies: A Retrospective Cohort Study. Acta Cytologica. 2024. doi:10.1159/000541374

BibTeX
@article{Verma2024,
  author = {Verma, Adarsh and McDowell, Rhonda and Porreca, Anthony},
  title = {Fine-Needle Aspiration versus the CytoCore ® Motorized Rotating Needle Device for Thyroid Nodule Biopsies: A Retrospective Cohort Study},
  journal = {Acta Cytologica},
  year = {2024},
  doi = {10.1159/000541374}
}
Kostenlos starten →