EUVIMEDEUVIMED entdecken
Crossref

In vivo Assessment of a Manual Single Lumen Alternating Micro-Batch Hemodiafiltration System

Sabrina L Lanker, Christopher J Pino, H David Humes, Lakhmir Chawla, Kimberly A Johnston

Blood Purification · 2024 · 53 (8) · 650-656

DOI: 10.1159/000538358 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Blood Purification, online erschienen am 27.03.2024. Verfasst von Sabrina L Lanker und 4 weiteren Autor:innen. Beteiligte Einrichtung: Innovative BioTherapies Inc, Ann Arbor, Michigan, USA. Hauptthemen laut MEDLINE: Hemodiafiltration. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 5 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The manual single lumen alternating micro-batch hemodiafiltration (mSLAMB) system is a closed-loop dialysis system designed to provide kidney support in emergency situations (e.g., fluid overload, hyperkalemia, acidemia). If done repeatedly in small batches and at high flow rates, this system was found to achieve clearance levels comparable to traditional renal replacement therapy.

Methods

Using a porcine model, uremic toxins and exogenous fluorescent tracer concentrations were successfully lowered within just 1 h of treatment.

Results

With a maximal dialysate flow, mSLAMB can achieve decreases in serum potassium concentration of >0.5 mmol/L/h. With the mSLAMB hemodiafiltration system, micro-batch processing was also successful in removing up to 250 mL of ultrafiltrate in 8 cycles.

Conclusion

This process could create a better fluid balance allowing for administering therapeutic fluids such as sodium bicarbonate in the clinic. Electrolyte imbalance and volume overload remain severe life-threatening emergencies in low resource settings; therefore, mSLAMB should be explored further due to its modest vascular access requirements, low cost, and ability to be performed without electricity or batteries.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemodiafiltrationAnimalsSwinePotassium

Stichwörter der Autor:innen

Acute kidney injuryAustere medical environmentLow-income countriesManual dialysisSLAMB

Substanzen

Potassium

Autor:innen und Institutionen

  • Innovative BioTherapies Inc, Ann Arbor, Michigan, USA.
  • Innovative BioTherapies Inc, Ann Arbor, Michigan, USA, cpino@innbio.com.
  • Veterans Affairs Medical Center, San Diego, California, USA.

Zitieren

APA

Lanker, S. L., Pino, C. J., Humes, H. D., Chawla, L., & Johnston, K. A. (2024). In vivo Assessment of a Manual Single Lumen Alternating Micro-Batch Hemodiafiltration System. Blood Purification, 53(8), 650-656. https://doi.org/10.1159/000538358

Vancouver

Lanker SL, Pino CJ, Humes HD, Chawla L, Johnston KA. In vivo Assessment of a Manual Single Lumen Alternating Micro-Batch Hemodiafiltration System. Blood Purification. 2024;53(8):650-656. doi:10.1159/000538358

BibTeX
@article{Lanker2024,
  author = {Lanker, Sabrina L and Pino, Christopher J and Humes, H David and Chawla, Lakhmir and Johnston, Kimberly A},
  title = {In vivo Assessment of a Manual Single Lumen Alternating Micro-Batch Hemodiafiltration System},
  journal = {Blood Purification},
  year = {2024},
  volume = {53},
  number = {8},
  pages = {650--656},
  doi = {10.1159/000538358}
}
Kostenlos starten →