EUVIMEDEUVIMED entdecken
Crossref

Compromised Accuracy of Stereotactic Target Delineation Associated with Computed Tomography-Based Frame Registration: A Comparative Analysis of Magnetic Resonance Imaging-Computed Tomography Fusion

Junhyung Kim, Sunwoo Jang, Seok Ho Hong, Sang Ryong Jeon

Stereotactic and Functional Neurosurgery · 2024

DOI: 10.1159/000534999 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Stereotactic and Functional Neurosurgery, 2024. Verfasst von Junhyung Kim und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurological Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea, jhkim2ns@gmail.com. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Surgery, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Recent advancements in stereotactic neurosurgical techniques have become increasingly reliant on image-based target planning. We devised a case-phantom comparative analysis to evaluate the target registration errors arising during the magnetic resonance imaging (MRI)-computed tomography (CT) image fusion process.

Methods

For subjects whose preoperative MRI and CT images both contained fiducial frame localizers, we investigated discrepancies in target coordinates derived from frame registration based on either MRI or CT. We generated a phantom target through an image fusion process, merging the framed CT images with their corresponding reference MRIs after masking their fiducial indicators. This phantom target was then compared with the original during each instance of target planning.

Results

In our investigative study with 26 frame registrations, a systematic error in the y-axis was observed as −0.89 ± 0.42 mm across cases using either conventional CT and/or cone-beam CT (O-arm). For the z-axis, errors varied on a case-by-case basis, recording at +0.64 ± 1.09 mm with a predominant occurrence in those merged with cone-beam CT. Collectively, these errors resulted in an average Euclidean error of 1.33 ± 0.93 mm.

Conclusion

Our findings suggest that the accuracy of frame-based stereotactic planning is potentially compromised during MRI-CT fusion process. Practitioners should recognize this issue, underscoring a pressing need for strategies and advancements to optimize the process.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedSurgery, Computer-AssistedHumansImaging, Three-DimensionalStereotaxic TechniquesMagnetic Resonance Imaging

Stichwörter der Autor:innen

Deep brain stimulation surgeryFrame-based stereotaxyImage registrationO-armTarget registration error

Autor:innen und Institutionen

  • Department of Neurological Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea, jhkim2ns@gmail.com.
  • Department of Neurological Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.

Zitieren

APA

Kim, J., Jang, S., Hong, S. H., & Jeon, S. R. (2024). Compromised Accuracy of Stereotactic Target Delineation Associated with Computed Tomography-Based Frame Registration: A Comparative Analysis of Magnetic Resonance Imaging-Computed Tomography Fusion. Stereotactic and Functional Neurosurgery. https://doi.org/10.1159/000534999

Vancouver

Kim J, Jang S, Hong SH, Jeon SR. Compromised Accuracy of Stereotactic Target Delineation Associated with Computed Tomography-Based Frame Registration: A Comparative Analysis of Magnetic Resonance Imaging-Computed Tomography Fusion. Stereotactic and Functional Neurosurgery. 2024. doi:10.1159/000534999

BibTeX
@article{Kim2024,
  author = {Kim, Junhyung and Jang, Sunwoo and Hong, Seok Ho and Jeon, Sang Ryong},
  title = {Compromised Accuracy of Stereotactic Target Delineation Associated with Computed Tomography-Based Frame Registration: A Comparative Analysis of Magnetic Resonance Imaging-Computed Tomography Fusion},
  journal = {Stereotactic and Functional Neurosurgery},
  year = {2024},
  doi = {10.1159/000534999}
}