Fachartikel in der Zeitschrift Stereotactic and Functional Neurosurgery, online erschienen am 07.02.2022. Verfasst von Lauren H Hammer und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurology, University of California, San Francisco, San Francisco, California, USA. Hauptthemen laut MEDLINE: Artifacts, Deep Brain Stimulation. Die Arbeit wurde bisher 66-mal zitiert (Crossref) und nennt 7 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NINDS NIH HHS.
Abstract (Originalsprache)
Background
The Medtronic “Percept” is the first FDA-approved deep brain stimulation (DBS) device with sensing capabilities during active stimulation. Its real-world signal-recording properties have yet to be fully described.
Objective
This study details three sources of artifact (and potential mitigations) in local field potential (LFP) signals collected by the Percept and assesses the potential impact of artifact on the future development of adaptive DBS (aDBS) using this device.
Methods
LFP signals were collected from 7 subjects in both experimental and clinical settings. The presence of artifacts and their effect on the spectral content of neural signals were evaluated in both the stimulation ON and OFF states using three distinct offline artifact removal techniques.
Results
Template subtraction successfully removed multiple sources of artifact, including (1) electrocardiogram (ECG), (2) nonphysiologic polyphasic artifacts, and (3) ramping-related artifacts seen when changing stimulation amplitudes. ECG removal from stimulation ON (at 0 mA) signals resulted in spectral shapes similar to OFF stimulation spectra (averaged difference in normalized power in theta, alpha, and beta bands ≤3.5%). ECG removal using singular value decomposition was similarly successful, though required subjective researcher input. QRS interpolation produced similar recovery of beta-band signal but resulted in residual low-frequency artifact.
Conclusions
Artifacts present when stimulation is enabled notably affected the spectral properties of sensed signals using the Percept. Multiple discrete artifacts could be successfully removed offline using an automated template subtraction method. The presence of unrejected artifact likely influences online power estimates, with the potential to affect aDBS algorithm performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurology, University of California, San Francisco, San Francisco, California, USA.
- Department of Neurological Surgery, University of California, San Francisco, San Francisco, California, USA.
Förderung
- NINDS NIH HHS
Zitieren
APA
Hammer, L. H., Kochanski, R. B., Starr, P. A., & Little, S. (2022). Artifact Characterization and a Multipurpose Template-Based Offline Removal Solution for a Sensing-Enabled Deep Brain Stimulation Device. Stereotactic and Functional Neurosurgery, 100(3), 168-183. https://doi.org/10.1159/000521431
Vancouver
Hammer LH, Kochanski RB, Starr PA, Little S. Artifact Characterization and a Multipurpose Template-Based Offline Removal Solution for a Sensing-Enabled Deep Brain Stimulation Device. Stereotactic and Functional Neurosurgery. 2022;100(3):168-183. doi:10.1159/000521431
BibTeX
@article{Hammer2022,
author = {Hammer, Lauren H and Kochanski, Ryan B and Starr, Philip A and Little, Simon},
title = {Artifact Characterization and a Multipurpose Template-Based Offline Removal Solution for a Sensing-Enabled Deep Brain Stimulation Device},
journal = {Stereotactic and Functional Neurosurgery},
year = {2022},
volume = {100},
number = {3},
pages = {168--183},
doi = {10.1159/000521431}
}
EUVIMED