EUVIMEDEUVIMED entdecken
Open Access

Vitrectomy-Assisted Biopsy: An in vitro Study on the Impact of Cut Rate and Probe Size

Erlend Ulltang, Jens Folke Kiilgaard, Nazanin Mola, David Scheie, Steffen Heegaard, Jørgen Krohn

Ocular Oncology and Pathology · 2021 · 7 (5) · 346-352

DOI: 10.1159/000516960 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ocular Oncology and Pathology, online erschienen am 08.07.2021. Verfasst von Erlend Ulltang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Haukeland University Hospital, Bergen, Norway. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study was to optimize the technique of performing vitrectomy-assisted biopsy of intraocular tumors by comparing the cytohistological findings in specimens obtained with different vitrectomy probes and cut rates.

Methods

Vitrectomy-assisted biopsies were taken from a fresh porcine liver. For each sampling, the vacuum level was 300 mm Hg. The following parameters were compared; cut rate (60, 600 and 6,000 cuts per minute [cpm]), probe type (standard and two-dimensional cutting [TDC]), and probe diameter (23-gauge and 25-gauge). The specimens were assessed by automated whole-slide imaging analysis and conventional light microscopy.

Results

Seventy-two biopsies were analyzed for the number of hepatocytes, total area of tissue fragments, and total stained area of each microscope slide. For all probe types, these parameters were significantly and positively correlated with the cut rate. TDC probes led to significantly higher scores than those of standard probes, independent of the cut rate. There were no significant differences in results when using 23-gauge or 25-gauge standard probes. Light microscopic examination demonstrated well-preserved cells sufficient for cytohistological analyses in all investigated cases.

Conclusions

The higher the cut rate, the larger is the amount of aspirated cellular material. There were no significant differences between 23-gauge and 25-gauge biopsies. Cut rates up to 6,000 cpm did not adversely affect the cytohistological features of the samples.

Schlagwörter

Stichwörter der Autor:innen

23-gauge vitrectomy25-gauge vitrectomyCut rateDigital image analysisPars plana vitrectomyVitrectomy-assisted biopsy

Autor:innen und Institutionen

  • Department of Ophthalmology, Haukeland University Hospital, Bergen, Norway.
  • Department of Ophthalmology, Copenhagen University Hospital, Copenhagen, Denmark.
  • Department of Pathology, Haukeland University Hospital, Bergen, Norway.
  • Section for Pathology, Department of Clinical Medicine, University of Bergen, Bergen, Norway.
  • Department of Pathology, Copenhagen University Hospital, Copenhagen, Denmark.
  • Section of Ophthalmology, Department of Clinical Medicine, University of Bergen, Bergen, Norway.

Zitieren

APA

Ulltang, E., Kiilgaard, J. F., Mola, N., Scheie, D., Heegaard, S., & Krohn, J. (2021). Vitrectomy-Assisted Biopsy: An in vitro Study on the Impact of Cut Rate and Probe Size. Ocular Oncology and Pathology, 7(5), 346-352. https://doi.org/10.1159/000516960

Vancouver

Ulltang E, Kiilgaard JF, Mola N, Scheie D, Heegaard S, Krohn J. Vitrectomy-Assisted Biopsy: An in vitro Study on the Impact of Cut Rate and Probe Size. Ocular Oncology and Pathology. 2021;7(5):346-352. doi:10.1159/000516960

BibTeX
@article{Ulltang2021,
  author = {Ulltang, Erlend and Kiilgaard, Jens Folke and Mola, Nazanin and Scheie, David and Heegaard, Steffen and Krohn, Jørgen},
  title = {Vitrectomy-Assisted Biopsy: An in vitro Study on the Impact of Cut Rate and Probe Size},
  journal = {Ocular Oncology and Pathology},
  year = {2021},
  volume = {7},
  number = {5},
  pages = {346--352},
  doi = {10.1159/000516960}
}
Kostenlos starten →