Taper Fracture of Uncemented Femoral Stem after Total Hip Arthroplasty
Case Reports in Orthopedic Research · 2018
DOI: 10.1159/000495359 ↗Kurz gefasst
Fachartikel in der Zeitschrift Case Reports in Orthopedic Research, 2018. Verfasst von Salim Al Habsi und 2 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Fracture through the tapered femoral neck of total hip arthroplasty is extremely rare. We report the taper neck fracture of a cementless stem in an extensively porous-coated femoral prosthesis used for total hip arthroplasty due to chronic hip infection 8 years previously in a 42-year-old man, weighing 110 kg with a body mass index of 39. We consider that cyclical loading stresses, fretting corrosion, and high body mass index are possible potential risk factors of such fracture.
Zitieren
APA
Habsi, S. A., Luthra, J. S., & Badei, H. A. (2018). Taper Fracture of Uncemented Femoral Stem after Total Hip Arthroplasty. Case Reports in Orthopedic Research. https://doi.org/10.1159/000495359
Vancouver
Habsi SA, Luthra JS, Badei HA. Taper Fracture of Uncemented Femoral Stem after Total Hip Arthroplasty. Case Reports in Orthopedic Research. 2018. doi:10.1159/000495359