Abstract (Originalsprache)
Intracochlear pressure changes during the cochlear implant insertion are assumed to be an important contributor to hearing preservation. The aim was to observe intracochlear pressure changes by different round window opening sizes and different hydrophilic electrode conditions. The experiments were performed in a cochlear model with a micropressure sensor in the helicotrema area. Different artificial round window membrane and different moisturized electrode conditions were compared. A punctured round window causes a significantly higher and an indirect moisturized electrode condition a significantly lower intracochlear pressure change. The degree of round window opening and the hydrophilic character of an electrode during insertion affect the intracochlear pressure significantly in a model.
Zitieren
APA
Todt, I., Ernst, A., & Mittmann, P. (2016). Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode. Audiology and Neurotology Extra. https://doi.org/10.1159/000442515
Vancouver
Todt I, Ernst A, Mittmann P. Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode. Audiology and Neurotology Extra. 2016. doi:10.1159/000442515
BibTeX
@article{Todt2016,
author = {Todt, Ingo and Ernst, Arneborg and Mittmann, Philipp},
title = {Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode},
journal = {Audiology and Neurotology Extra},
year = {2016},
doi = {10.1159/000442515}
}
EUVIMED