EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode

Ingo Todt, Arneborg Ernst, Philipp Mittmann

Audiology and Neurotology Extra · 2016

DOI: 10.1159/000442515 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Audiology and Neurotology Extra, 2016. Verfasst von Ingo Todt und 2 weiteren Autor:innen. Die Arbeit wurde bisher 23-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Intracochlear pressure changes during the cochlear implant insertion are assumed to be an important contributor to hearing preservation. The aim was to observe intracochlear pressure changes by different round window opening sizes and different hydrophilic electrode conditions. The experiments were performed in a cochlear model with a micropressure sensor in the helicotrema area. Different artificial round window membrane and different moisturized electrode conditions were compared. A punctured round window causes a significantly higher and an indirect moisturized electrode condition a significantly lower intracochlear pressure change. The degree of round window opening and the hydrophilic character of an electrode during insertion affect the intracochlear pressure significantly in a model.

Zitieren

APA

Todt, I., Ernst, A., & Mittmann, P. (2016). Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode. Audiology and Neurotology Extra. https://doi.org/10.1159/000442515

Vancouver

Todt I, Ernst A, Mittmann P. Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode. Audiology and Neurotology Extra. 2016. doi:10.1159/000442515

BibTeX
@article{Todt2016,
  author = {Todt, Ingo and Ernst, Arneborg and Mittmann, Philipp},
  title = {Effects of Round Window Opening Size and Moisturized Electrodes on Intracochlear Pressure Related to the Insertion of a Cochlear Implant Electrode},
  journal = {Audiology and Neurotology Extra},
  year = {2016},
  doi = {10.1159/000442515}
}
Kostenlos starten →