Accuracy and Precision of Third‐Generation Tympanic Thermometers With Varying Calibration Intervals: A Multicenter Cross‐Sectional Study
Nursing Research and Practice · 2026-01 · 2026 (1)
DOI: 10.1155/nrp/8453356 ↗Kurz gefasst
Fachartikel in der Zeitschrift Nursing Research and Practice, online erschienen am 11.03.2026. Verfasst von Mattia Morri und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nursing, Technical and Rehabilitation Care, IRCCS Rizzoli Orthopedic Institute, Bologna, Italy. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aim
To evaluate intra‐ and interoperator reliability and accuracy of the third‐generation tympanic thermometer (GeniusTM3) and to explore the relationship between measurement error and calibration intervals of the tympanic thermometer.
Design
Open‐label, pragmatic multicenter cross‐sectional study.
Methods
The study was carried out in intensive care units and operating theaters across five hospitals in Northern Italy. Patients of any age admitted to intensive care units or undergoing surgical procedures in operating theaters between December 2023 and October 2024 were enrolled. Two trained nurses independently took the temperature three consecutive times using the same tympanic thermometer. Core temperature was measured using one of the gold standard methods. The measurement error was calculated as the difference between the gold standard measurement and the measurement made with the tympanic temperature.
Results
A total of 550 patients were enrolled. The precision of the tympanic thermometer was shown to be excellent in both intraobserver and interobserver reliability. The tympanic thermometer consistently underestimated body temperature, with a mean measurement error of 0.3°C and a limits of agreement of ±1.3°C. The generalized linear model (GLM) showed calibration interval, recruiting hospital, and the variable defined by their interaction as factors that were independently and statistically significant predictors of accuracy variation.
Conclusions
The third‐generation tympanic thermometer (GeniusTM3) represents a reliable tool for ensuring high accuracy in terms of both intraoperator and interoperator concordance. However, the accuracy of the device has been significantly affected by the timing of the calibration: a calibration interval of more than 6 months results in a significant deterioration in accuracy.
Autor:innen und Institutionen
- Department of Nursing, Technical and Rehabilitation Care, IRCCS Rizzoli Orthopedic Institute, Bologna, Italy.
- Center for Medical Sciences (CISMed), University of Trento, Trento, Italy, unitn.it.
- Medical Genetics Unit, IRCCS Azienda Ospedaliero-Universitaria di Bologna, Bologna, Italy.
- Department of Research and Innovation (DAIRI) - Azienda Ospedaliero- Universitaria SS Antonio e Biagio e Cesare Arrigo, Alessandria, Italy.
- Health Professions Unit, Integrated University Hospital of Verona, Verona, Italy.
- Department of Healthcare Professionals, Meyer Children's Hospital - IRCCS, Florence, Italy.
ORCID-Profile: Mattia Morri, Anna Brugnolli, Lea Godino, Tatiana Bolgeo, Domenica Gazineo, Marina Maffeo, Cristiana Forni
Zitieren
APA
Morri, M., Brugnolli, A., Godino, L., Bolgeo, T., Azzolini, M., Ciofi, D., Gardalini, M., Grezzani, C., Bettini, K., Gazineo, D., Maffeo, M., & Forni, C. (2026). Accuracy and Precision of Third‐Generation Tympanic Thermometers With Varying Calibration Intervals: A Multicenter Cross‐Sectional Study. Nursing Research and Practice, 2026(1). https://doi.org/10.1155/nrp/8453356
Vancouver
Morri M, Brugnolli A, Godino L, Bolgeo T, Azzolini M, Ciofi D, et al. Accuracy and Precision of Third‐Generation Tympanic Thermometers With Varying Calibration Intervals: A Multicenter Cross‐Sectional Study. Nursing Research and Practice. 2026;2026(1). doi:10.1155/nrp/8453356
BibTeX
@article{Morri2026,
author = {Morri, Mattia and Brugnolli, Anna and Godino, Lea and Bolgeo, Tatiana and Azzolini, Marianna and Ciofi, Daniele and Gardalini, Menada and Grezzani, Claudia and Bettini, Katia and Gazineo, Domenica and Maffeo, Marina and Forni, Cristiana},
title = {Accuracy and Precision of Third‐Generation Tympanic Thermometers With Varying Calibration Intervals: A Multicenter Cross‐Sectional Study},
journal = {Nursing Research and Practice},
year = {2026},
volume = {2026},
number = {1},
doi = {10.1155/nrp/8453356}
}