EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Near‐Infrared Spectroscopy Sensors Placed Frontally and Parietally to Detect Large Vessel Occlusion Stroke

Frontale und parietale Nahinfrarotspektroskopie-Sensoren zum Nachweis eines großen Gefäßverschluss-Schlaganfalls

Joonas Kiili iD, Pauli Vuorinen iD, Jyrki Ollikainen iD, Pasi Jolma iD, Heini Huhtala iD, Piritta Setälä iD, Sanna Hoppu iD

Neurology Research International · 2026-01 · 2026 (1)

DOI: 10.1155/nri/7575719 ↗

Auf einen Blick

Bei 6 CTA-bestätigten LVO-Schlaganfallpatient:innen zeigte NIRS weder frontal noch parietal verlässliche Unterschiede der rSO2. Die Methode erschien dafür ungeeignet.

Fragestellung

Ist zwei-kanalige NIRS mit frontaler oder parietaler Sondenplatzierung für die Erkennung eines anterioren LVO geeignet, und macht die Platzierung einen Unterschied?

Methode

Analyse von NIRS-Messungen bei 6 CTA-bestätigten LVO-Schlaganfallpatient:innen; 4 Sensoren pro Kopf, frontal und parietal platziert; Korrelation mit perfusionsbildgebenden Veränderungen.

Ergebnisse

Frontale Sensoren: rSO2 war 1–8 Prozentpunkte vor und 0–9 Prozentpunkte nach EVT auf der betroffenen Seite höher; parietale Sensoren: vor EVT 3–9, nach EVT meist 3–7 Prozentpunkte höher auf der gesunden Seite. Die rSO2 änderte sich nicht signifikant.

Einschränkungen

Bei 1 Patient:in keine frontalen Werte, bei 1 Patient:in keine parietalen Werte verfügbar.

Schlussfolgerung der Autor:innen

Weder frontale noch parietale NIRS-Sensorplatzierung detektierte Unterschiede der rSO2 bei LVO; NIRS erscheint daher als diagnostisches Verfahren für LVO ungeeignet.

Kernaussagen

  • 6 CTA-bestätigte LVO-Schlaganfallpatient:innen
  • Keine signifikanten rSO2-Unterschiede zwischen betroffener und gesunder Seite
  • NIRS erschien als Diagnostik für LVO ungeeignet

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Neurology Research International, online erschienen am 09.11.2026. Verfasst von Joonas Kiili und 6 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine and Health Technology Tampere University Tampere Finland uta.fi. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

We aimed to determine whether two‐channelled near‐infrared spectroscopy (NIRS) was feasible for recognising anterior large vessel occlusion (LVO) with a device suitable for prehospital use and whether probe placement (frontal vs. parietal) made a difference.

Methods

We analysed NIRS readings of six computed tomography angiography (CTA)‐confirmed LVO‐stroke patients at Tampere University Hospital during the six‐month study period. We placed four sensors on patients’ heads: two frontally and two parietally. We provided a detailed description and analysis of these recordings and their correlation with imaging‐confirmed changes in perfusion.

Results

In the frontally placed sensors, the rSO 2 was 1–8 percentage points higher on the affected side before and 0–9 percentage points higher on the affected side after endovascular treatment (EVT). rSO2 values did not change significantly. One patient showed higher rSO 2 on the healthy side (5 percentage points) before and after EVT. We could not recover frontal values from one patient. In parietally placed censors, all patients had 3–9 percentage points higher rSO 2 on the healthy side before EVT. After EVT, 3 patients had 3–7 percentage points higher rSO 2 on the healthy side. In two cases, after EVT, the difference in rSO 2 changed to 1‐2 percentage points higher on the affected side. We could not record parietal readings in one patient.

Conclusions

Neither NIRS sensor placement detected differences in rSO 2 in LVO; therefore, NIRS appears unsuitable as a diagnostic tool for LVO. Future research should focus on other technologies.

Schlagwörter

Stichwörter der Autor:innen

NIRSlarge vessel occlusionnear-infrared spectroscopystrokethrombectomy

Autor:innen und Institutionen

  • Faculty of Medicine and Health Technology Tampere University Tampere Finland uta.fi.
  • Emergency Medical Services Centre for Prehospital Emergency Care Wellbeing Services County of Pirkanmaa Tampere Finland.
  • Department of Anaesthesiology Hospital Nova Wellbeing Services County of Central Finland Jyväskylä Finland.
  • Department of Neurosciences and Rehabilitation Tampere University Hospital Tampere Finland pshp.fi.
  • Faculty of Social Sciences Tampere University Tampere Finland uta.fi.

ORCID-Profile: Joonas Kiili, Pauli Vuorinen, Jyrki Ollikainen, Pasi Jolma, Heini Huhtala, Piritta Setälä, Sanna Hoppu

Zitieren

APA

Kiili, J., Vuorinen, P., Ollikainen, J., Jolma, P., Huhtala, H., Setälä, P., & Hoppu, S. (2026). Near‐Infrared Spectroscopy Sensors Placed Frontally and Parietally to Detect Large Vessel Occlusion Stroke. Neurology Research International, 2026(1). https://doi.org/10.1155/nri/7575719

Vancouver

Kiili J, Vuorinen P, Ollikainen J, Jolma P, Huhtala H, Setälä P, et al. Near‐Infrared Spectroscopy Sensors Placed Frontally and Parietally to Detect Large Vessel Occlusion Stroke. Neurology Research International. 2026;2026(1). doi:10.1155/nri/7575719

BibTeX
@article{Kiili2026,
  author = {Kiili, Joonas and Vuorinen, Pauli and Ollikainen, Jyrki and Jolma, Pasi and Huhtala, Heini and Setälä, Piritta and Hoppu, Sanna},
  title = {Near‐Infrared Spectroscopy Sensors Placed Frontally and Parietally to Detect Large Vessel Occlusion Stroke},
  journal = {Neurology Research International},
  year = {2026},
  volume = {2026},
  number = {1},
  doi = {10.1155/nri/7575719}
}
Kostenlos starten →