Precision of Pachymetry Measured With a Portable High‐Resolution Ultrasound, a Standard Ultrasound, and Two Scheimpflug Pachymeters

Anqi Liu, Youhan Ao, Yanyan Zhang, Zequan Xu, Feng Liu, Lina Mei, Mei Ge, Yifei Huang iD, Liqiang Wang iD

Journal of Ophthalmology · 2026-01 · 2026 (1)

DOI: 10.1155/joph/4314297 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Ophthalmology, online erschienen am 28.04.2026. Verfasst von Anqi Liu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Chinese PLA General Hospital, No. 28 Fuxing Road Haidian District, Beijing, 100853, China, 301hospital.com.cn. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Subject To assess the precision (repeatability and reproducibility) and agreement of central corneal thickness (CCT) measurements by a high‐resolution ultrasound (E‐pach), a standard ultrasound, and two Scheimpflug pachymeters in healthy eyes.

Methods

Instrument diagnostic test study. A total of 100 healthy volunteers were enrolled to measure right‐eye CCT by a high‐resolution ultrasound (E‐pach), a standard ultrasound (A‐scan device), and two Scheimpflug pachymeters (Pentacam and Corvis ST). To assess repeatability and reproducibility, the test−retest repeatability (TRT) and intraclass correlation coefficient (ICC) were calculated. The agreement among the four devices was evaluated with Bland−Altman plots.

Results

The high‐resolution ultrasound (E‐pach) showed repeatability (ICC = 0.9981), interobserver reproducibility (ICC = 0.9971), and intersession reproducibility (ICC = 0.9825); the standard ultrasound and two Scheimpflug pachymeters also showed similar repeatability (all ICC ≥ 0.9679), interobserver reproducibility (all ICC ≥ 0.9730), and intersession reproducibility (all ICC ≥ 0.9647). However, the high‐resolution ultrasound yielded CCT values that were obviously lower than those of the standard ultrasound and Pentacam ( p < 0.001) but higher than those of the Corvis ST ( p < 0.001). The 95% limits of agreement (LoA) in the Bland−Altman plots were 44.5 µm (high‐resolution ultrasound vs. standard ultrasound), 34.9 µm (high‐resolution ultrasound vs. Corvis CT), and 32.5 µm (high‐resolution ultrasound vs. Pentacam).

Conclusions

The high‐resolution ultrasound is a portable, reliable, and inexpensive pachymeter. However, the CCT values obtained from the high‐resolution ultrasound are not interchangeable with those from Pentacam, Corvis ST, and standard ultrasound.

Schlagwörter

Stichwörter der Autor:innen

agreementcentral corneal thicknesshigh-resolution ultrasoundpachymetryrepeatabilityreproducibility

Autor:innen und Institutionen

  • Department of Ophthalmology, Chinese PLA General Hospital, No. 28 Fuxing Road Haidian District, Beijing, 100853, China, 301hospital.com.cn.

ORCID-Profile: Yifei Huang, Liqiang Wang

Zitieren

APA

Liu, A., Ao, Y., Zhang, Y., Xu, Z., Liu, F., Mei, L., Ge, M., Huang, Y., & Wang, L. (2026). Precision of Pachymetry Measured With a Portable High‐Resolution Ultrasound, a Standard Ultrasound, and Two Scheimpflug Pachymeters. Journal of Ophthalmology, 2026(1). https://doi.org/10.1155/joph/4314297

Vancouver

Liu A, Ao Y, Zhang Y, Xu Z, Liu F, Mei L, et al. Precision of Pachymetry Measured With a Portable High‐Resolution Ultrasound, a Standard Ultrasound, and Two Scheimpflug Pachymeters. Journal of Ophthalmology. 2026;2026(1). doi:10.1155/joph/4314297

BibTeX
@article{Liu2026,
  author = {Liu, Anqi and Ao, Youhan and Zhang, Yanyan and Xu, Zequan and Liu, Feng and Mei, Lina and Ge, Mei and Huang, Yifei and Wang, Liqiang},
  title = {Precision of Pachymetry Measured With a Portable High‐Resolution Ultrasound, a Standard Ultrasound, and Two Scheimpflug Pachymeters},
  journal = {Journal of Ophthalmology},
  year = {2026},
  volume = {2026},
  number = {1},
  doi = {10.1155/joph/4314297}
}
Kostenlos starten →