EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Developing and Validating a Robust RP‐HPLC Method for Metoclopramide and Camylofin Simultaneous Analysis Using Response Surface Methodology

Ahmed Hussain Jawhari iD, Zeinhom H Mohamed iD

International Journal of Analytical Chemistry · 2025-01 · 2025 (1)

DOI: 10.1155/ianc/5543392 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Analytical Chemistry, online erschienen am 15.10.2025. Verfasst von Ahmed Hussain Jawhari und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physical Sciences, Chemistry Division, College of Science, Jazan University, P.O. Box. 114, Jazan 45142, Saudi Arabia. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Jazan University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this study, the establishment and validation of a stable reversed‐phase high‐performance liquid chromatography (RP‐HPLC) method for the concomitant estimation of the two drugs in dosage forms are presented. Method optimization was achieved by response surface methodology (RSM) using Design Expert Software 13, taking into account the special physicochemical characteristics of metoclopramide (MET) (a moderately polar molecule, pKa 9.5) and camylofin (CAM) (a less polar, hydrophobic molecule, pKa 8.7). Chromatographic resolution was achieved on a phenyl‐hexyl column under isocratic mobile phase mode in which methanol and 20 mM ammonium acetate buffer (pH 3.5) were used to provide maximum analyte interaction and resolution. The method was found to have good linearity for both analytes ( R 2 > 0.999) over the concentration ranges studied. Limits of detection were 0.23 and 0.15 μg/mL for MET and CAM, respectively, and corresponding limits of quantification were 0.35 and 0.42 μg/mL, respectively. Recovery tests gave high precision values of 98.2%–101.5%, while intra‐ and inter‐day precision in relative standard deviation (RSD) was below 2%. The method was effectively applied for the analysis of commercial tablet formulations, confirming its reliability and suitability for routine quality control and regulatory analyses. Overall, the validated RP‐HPLC method provides a sensitive, accurate, and efficient means of simultaneous determination of MET and CAM in pharmaceutical dosage forms.

Schlagwörter

Stichwörter der Autor:innen

HPLC method developmentcamylofinchromatographic separationmetoclopramideresponse surface methodology

Autor:innen und Institutionen

  • Department of Physical Sciences, Chemistry Division, College of Science, Jazan University, P.O. Box. 114, Jazan 45142, Saudi Arabia.

ORCID-Profile: Ahmed Hussain Jawhari, Zeinhom H Mohamed

Förderung

  • Jazan University

Zitieren

APA

Jawhari, A. H., & Mohamed, Z. H. (2025). Developing and Validating a Robust RP‐HPLC Method for Metoclopramide and Camylofin Simultaneous Analysis Using Response Surface Methodology. International Journal of Analytical Chemistry, 2025(1). https://doi.org/10.1155/ianc/5543392

Vancouver

Jawhari AH, Mohamed ZH. Developing and Validating a Robust RP‐HPLC Method for Metoclopramide and Camylofin Simultaneous Analysis Using Response Surface Methodology. International Journal of Analytical Chemistry. 2025;2025(1). doi:10.1155/ianc/5543392

BibTeX
@article{Jawhari2025,
  author = {Jawhari, Ahmed Hussain and Mohamed, Zeinhom H},
  title = {Developing and Validating a Robust RP‐HPLC Method for Metoclopramide and Camylofin Simultaneous Analysis Using Response Surface Methodology},
  journal = {International Journal of Analytical Chemistry},
  year = {2025},
  volume = {2025},
  number = {1},
  doi = {10.1155/ianc/5543392}
}
Kostenlos starten →