Underrecognized Interaction Between PICO Dressings and Pacemakers: A Case From the ICU
Case Reports in Anesthesiology · 2026-01 · 2026 (1)
DOI: 10.1155/cria/2189554 ↗Kurz gefasst
Fachartikel in der Zeitschrift Case Reports in Anesthesiology, online erschienen am 04.08.2026. Verfasst von Reginald Caldecott und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesia and Intensive Care Medicine, University Hospital Galway, Galway, Ireland, hse.ie. Die Arbeit nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Modern pacemakers incorporate a magnet‐responsive switch that, on exposure to a sufficient static magnetic field, converts the device to fixed‐rate asynchronous (magnet‐mode) pacing. This is designed behaviour rather than malfunction. Interactions between this feature and magnet‐containing medical equipment are underrecognised. The PICO 7 negative pressure wound therapy (NPWT) system contains a magnet in its pump unit, which the manufacturer states can interfere with nearby pacemakers. Case Presentation An 81‐year‐old man was admitted to the High Dependency Unit after right hemicolectomy. He had atrial fibrillation with a slow ventricular response and a permanent pacemaker programmed VVI at 60 beats per minute. Overnight, telemetry showed a sustained paced rate of 100 beats per minute; he was asymptomatic. The PICO 7 pump unit was found resting on his chest, over the pacemaker generator. Repositioning the pump to the foot of the bed returned the paced rate to 60 beats per minute immediately, and device interrogation the following day confirmed normal pacemaker function.
Discussion
The fixed rate of 100 beats per minute, its dependence on pump position and immediate resolution on removal are characteristic of magnet‐mode activation rather than generic electromagnetic interference. Although the patient was asymptomatic, a sustained rate of 100 beats per minute carried a risk of myocardial ischaemia and heart failure decompensation given his cardiac history. Recognition and a simple bedside intervention resolved the episode.
Conclusion
Magnet‐containing medical equipment placed near a cardiac implantable electronic device can trigger magnet‐mode pacing. The risk depends on the position of the magnet‐bearing component, not on the wound site. An unexplained change in a paced rate should prompt early consideration of a nearby magnetic source, because the remedy may be immediate and at the bedside.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anaesthesia and Intensive Care Medicine, University Hospital Galway, Galway, Ireland, hse.ie.
- Department of Anaesthesia, School of Medicine, University of Galway, Galway, Ireland, universityofgalway.ie.
ORCID-Profile: Reginald Caldecott, John Laffey
Zitieren
APA
Caldecott, R., & Laffey, J. (2026). Underrecognized Interaction Between PICO Dressings and Pacemakers: A Case From the ICU. Case Reports in Anesthesiology, 2026(1). https://doi.org/10.1155/cria/2189554
Vancouver
Caldecott R, Laffey J. Underrecognized Interaction Between PICO Dressings and Pacemakers: A Case From the ICU. Case Reports in Anesthesiology. 2026;2026(1). doi:10.1155/cria/2189554
BibTeX
@article{Caldecott2026,
author = {Caldecott, Reginald and Laffey, John},
title = {Underrecognized Interaction Between PICO Dressings and Pacemakers: A Case From the ICU},
journal = {Case Reports in Anesthesiology},
year = {2026},
volume = {2026},
number = {1},
doi = {10.1155/cria/2189554}
}
EUVIMED