Evaluation of Below‐Knee Prosthetic Socket Mechanical Characteristics via Experimental and Numerical Methods
Advances in Polymer Technology · 2026-01 · 2026 (1)
DOI: 10.1155/adv/6352808 ↗Kurz gefasst
Fachartikel in der Zeitschrift Advances in Polymer Technology, online erschienen am 22.11.2026. Verfasst von Saif M. Abbas und 2 weiteren Autor:innen. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This study aimed at assessing the mechanical and biomechanical behavior of multilayer prosthetic sockets determined for transtibial amputation made up of polymer, carbon, and glass fibers. Measurements in one 45‐year‐old subject (83 kg, height: 167 cm) were evaluated, such as stump–socket interface pressure and strategy of composite mechanical properties. Two glass‐fiber layers increased the modulus of elasticity, yield strength, and ultimate tensile strength by 86.7%, 57.7%, and 32.2%, respectively, while two carbon‐fiber layers gave the following improvements of 51.85%, 89.3% and 64.3% as compared with group A (perlon only). The pressure test located on four sides (anterior, lateral, posterior, and medial) of the prosthetic socket. Peak pressure areas were located at the posterior (185 kPa) and lateral (160 kPa) aspects, highlighting the necessity to avoid anterior loading during score formation to protect the tibia and fibula. Results from static safety factor analysis were 0.85, 1.7, and 2.127 for Groups A (perlon), B (perlon and Glass fiber), and C (perlon and carbon fiber), respectively, with both Group B and SGR exceeding the commonly accepted biological safety limit. The maximum patellar tendon pressure was 125 kPa, and the lowest safety factor value of 2.8 in the anterodistal region showed that it had enough stability for being a structure. Close correspondence between finite element analysis (FEA) and experimental data (≤10% deviation) is indicative of the accuracy and validity of the computational model for comfort‐based design exploration. In summary, the prosthetic socket has acceptable structural integrity and user safety and function stability under physiological loading conditions.
Zitieren
APA
Abbas, S. M., Hamood, S. A., & Abdulaal, A. H. (2026). Evaluation of Below‐Knee Prosthetic Socket Mechanical Characteristics via Experimental and Numerical Methods. Advances in Polymer Technology, 2026(1). https://doi.org/10.1155/adv/6352808
Vancouver
Abbas SM, Hamood SA, Abdulaal AH. Evaluation of Below‐Knee Prosthetic Socket Mechanical Characteristics via Experimental and Numerical Methods. Advances in Polymer Technology. 2026;2026(1). doi:10.1155/adv/6352808
BibTeX
@article{Abbas2026,
author = {Abbas, Saif M. and Hamood, Sarah A. and Abdulaal, Alaa Hussein},
title = {Evaluation of Below‐Knee Prosthetic Socket Mechanical Characteristics via Experimental and Numerical Methods},
journal = {Advances in Polymer Technology},
year = {2026},
volume = {2026},
number = {1},
doi = {10.1155/adv/6352808}
}
EUVIMED