Validation on Method of Measuring the Knee Joint Force Sense and Application in Patients With Knee Osteoarthritis
Applied Bionics and Biomechanics · 2025-01 · 2025 (1)
DOI: 10.1155/abb/8851877 ↗Kurz gefasst
Fachartikel in der Zeitschrift Applied Bionics and Biomechanics, online erschienen am 03.07.2025. Verfasst von Zhixin Jin und 5 weiteren Autor:innen. Beteiligte Einrichtung: Beijing Sport University, Beijing 100084, China. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Beijing Municipal Natural Science Foundation.
Abstract (Originalsprache)
Background
Force sense reflects the efferent activity capacity of proprioception. Currently, methods for measuring force sense in the knee joint is lack. This study aims to validate a constructed test system and to explore the characteristics of force sense in patients with knee osteoarthritis (KOA).
Methods
Seventy‐five subjects were recruited: 30 for verifying the reliability of test system and exploring the impact of body mass index (BMI) and the striking force on force sense; 20 healthy adults and 25 KOA patients for the method application. The force sense test system comprised a self‐designed force application apparatus and a wireless surface electromyography (sEMG) device (DELSYS Inc, USA). The reflex contraction latency of muscles was considered as the force sense. The intraclass correlation coefficient (ICC) was used to verify the reliability.
Results
The reflex contraction latencies of biceps femoris (BF) was the shortest and ICC in the two tests is 0.950 ( p < 0.001). No significant differences in force sense were found between different BMI groups ( p = 0.065), and there was no notable interaction between BMI and striking force ( p = 0.283). A significant difference in force sense was observed between different striking forces ( p < 0.001). There were no significant differences in force sense of bilateral sides between healthy people and KOA patients ( p = 0.126, p = 0.315).
Conclusion
The force sense testing method established in this study is applicable for measuring knee joint force sense. BMI did not affect knee joint force sense but striking force did, and subjects with different BMI chose the same striking force; KOA did not influence the force sense of knee joint.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Beijing Sport University, Beijing 100084, China.
- School of Biomedical Engineering, Capital Medical University, Beijing 100069, China.
- Department of Orthopedics, Beijing Jishuitan Hospital, Capital Medical University, Beijing 100035, China.
ORCID-Profile: Kuan Zhang, Songhua Yan
Förderung
- Beijing Municipal Natural Science Foundation
Zitieren
APA
Jin, Z., Shi, J., Zhang, H., Zhou, X., Zhang, K., & Yan, S. (2025). Validation on Method of Measuring the Knee Joint Force Sense and Application in Patients With Knee Osteoarthritis. Applied Bionics and Biomechanics, 2025(1). https://doi.org/10.1155/abb/8851877
Vancouver
Jin Z, Shi J, Zhang H, Zhou X, Zhang K, Yan S. Validation on Method of Measuring the Knee Joint Force Sense and Application in Patients With Knee Osteoarthritis. Applied Bionics and Biomechanics. 2025;2025(1). doi:10.1155/abb/8851877
BibTeX
@article{Jin2025,
author = {Jin, Zhixin and Shi, Jingnan and Zhang, Haohua and Zhou, Xinglong and Zhang, Kuan and Yan, Songhua},
title = {Validation on Method of Measuring the Knee Joint Force Sense and Application in Patients With Knee Osteoarthritis},
journal = {Applied Bionics and Biomechanics},
year = {2025},
volume = {2025},
number = {1},
doi = {10.1155/abb/8851877}
}
EUVIMED