DYNEELAX Robotic Arthrometer Reliability and Feasibility on Healthy and Anterior Cruciate Ligament Injured/Reconstructed Persons
Translational Sports Medicine · 15.04.2024 · 2024 · 1-5
DOI: 10.1155/2024/3413466 ↗Kurz gefasst
Fachartikel in der Zeitschrift Translational Sports Medicine, erschienen am 15.04.2024. Verfasst von Nuno Nascimento und 8 weiteren Autor:innen. Beteiligte Einrichtung: Aspetar Orthopaedic and Sports Medicine Hospital, Sports City Street, P.O. Box 29222, Doha, Qatar. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Anterior cruciate ligament (ACL) injuries are commonly assessed using clinical examination and magnetic resonance imaging, but these methods have limitations in reproducibility and quantification. Instrumented laxity measurements using devices, like the DYNEELAX®, offer an alternative approach. However, to date, there is no human data on the DYNEELAX® and the reliability of these devices remains a subject of debate, and there is no consensus on appropriate knee tightening levels for testing. We hypothesized that the DYNEELAX®, with standardized knee tightening, would provide reliable measurements of knee laxity in adult volunteers.
Methods
This prospective cohort study involved 48 pain-free adult volunteers. Laxity measurements were taken using a robotic-type motorized instrument (DYNEELAX®) on two separate occasions, at least 1 h and no more than 8 h apart, with knee tightening forces of 90 N ± 5 N. Metrics of anterior tibial translation and internal/external tibial axial rotations were recorded.
Results
The device displayed excellent intrarater reliability for all the metrics, with intraclass correlation coefficients ranging from 0.91 to 0.96. Anterior translation exhibited the highest reliability (intraclass correlation coefficient = 0.96), with a minimum detectable change of 0.83 mm.
Conclusions
DYNEELAX® is reliable in measuring knee laxity in adult volunteers when using standardized stabilizing knee tightening forces of 90 ± 5 N. The most sensitive measurement parameters (in terms of minimum detectable change as a proportion of the observed range) were anterior translation (in mm) at 150 N and secondary compliance.
Autor:innen und Institutionen
- Aspetar Orthopaedic and Sports Medicine Hospital, Sports City Street, P.O. Box 29222, Doha, Qatar.
ORCID-Profile: Nuno Nascimento, Roula Kotsifaki, Emmanouil Papakostas, Elisabet Hagert, Rod Whiteley
Zitieren
APA
Nascimento, N., Kotsifaki, R., Papakostas, E., Zikria, B. A., Alkhelaifi, K., Hagert, E., Olory, B., D'Hooghe, P., & Whiteley, R. (2024). DYNEELAX Robotic Arthrometer Reliability and Feasibility on Healthy and Anterior Cruciate Ligament Injured/Reconstructed Persons. Translational Sports Medicine, 2024, 1-5. https://doi.org/10.1155/2024/3413466
Vancouver
Nascimento N, Kotsifaki R, Papakostas E, Zikria BA, Alkhelaifi K, Hagert E, et al. DYNEELAX Robotic Arthrometer Reliability and Feasibility on Healthy and Anterior Cruciate Ligament Injured/Reconstructed Persons. Translational Sports Medicine. 2024;2024:1-5. doi:10.1155/2024/3413466
BibTeX
@article{Nascimento2024,
author = {Nascimento, Nuno and Kotsifaki, Roula and Papakostas, Emmanouil and Zikria, Bashir A and Alkhelaifi, Khalid and Hagert, Elisabet and Olory, Bruno and D'Hooghe, Pieter and Whiteley, Rod},
title = {DYNEELAX Robotic Arthrometer Reliability and Feasibility on Healthy and Anterior Cruciate Ligament Injured/Reconstructed Persons},
journal = {Translational Sports Medicine},
year = {2024},
volume = {2024},
pages = {1--5},
doi = {10.1155/2024/3413466}
}
EUVIMED