Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study

Thanyawee Pengpan iD, Natch Rattanarungruangchai iD, Juthathip Dechjaithat, Phawinee Panthim, Puntarika Siricharuwong, Ausanai Prapan iD

Radiology Research and Practice · 16.04.2022 · 2022 · 1-10

DOI: 10.1155/2022/3482458 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiology Research and Practice, erschienen am 16.04.2022. Verfasst von Thanyawee Pengpan und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok 65000, Thailand. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Thailand Institute of Nuclear Technology.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to identify proper exposure techniques to maintain optimal diagnostic image quality with minimum radiation dose for anteroposterior chest X-ray projection in pediatric patients.

Methods

Briefly, an in-house developed pediatric chest phantom was constructed. Next, nanodot OSLDs were used for organ absorbed dose measurement and placed in the lung area, and the phantom was exposed to various exposure techniques (ranging from 50 to 70 kVp with 1.6, 2, and 2.5 mAs). After that, the phantom was used to assess image quality parameters, including SNR and CNR. Two radiologists assessed the subjective image quality using a visual grading analysis (VGA) technique. Finally, the figure of merit (FOM) was analyzed.

Results

The developed phantom was constructed successfully and could be useful for dose measurement and image quality assessment. The absorbed dose varied from 0.009 to 0.031 mGy for the range of exposure techniques used. SNR and CNR showed a gradually increasing trend, while kVp and mAs values were increased. The highest kVp (70 kVp) produced the highest SNR and CNR, exhibiting a significant difference compared with 50 and 60 kVp ( P < 0.05 ). The overall VGA score was 3.2 ± 0.3, and the low kVp technique demonstrated better image quality compared with the reference image.

Conclusion

The optimized exposure technique was identified as 60 kV and 2.5 mAs, indicating the highest FOM score. This work revealed practicable techniques that could be implemented into clinical practice for performing pediatric chest radiography.

Autor:innen und Institutionen

  • Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok 65000, Thailand.
  • Department of Radiation Dose Measurement and Assessment, Nuclear Technology Service Center, Thailand Institute of Nuclear Technology (Public Organization), Bangkok, 26120, Thailand.

ORCID-Profile: Thanyawee Pengpan, Natch Rattanarungruangchai, Ausanai Prapan

Förderung

  • Thailand Institute of Nuclear Technology

Zitieren

APA

Pengpan, T., Rattanarungruangchai, N., Dechjaithat, J., Panthim, P., Siricharuwong, P., & Prapan, A. (2022). Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study. Radiology Research and Practice, 2022, 1-10. https://doi.org/10.1155/2022/3482458

Vancouver

Pengpan T, Rattanarungruangchai N, Dechjaithat J, Panthim P, Siricharuwong P, Prapan A. Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study. Radiology Research and Practice. 2022;2022:1-10. doi:10.1155/2022/3482458

BibTeX
@article{Pengpan2022,
  author = {Pengpan, Thanyawee and Rattanarungruangchai, Natch and Dechjaithat, Juthathip and Panthim, Phawinee and Siricharuwong, Puntarika and Prapan, Ausanai},
  title = {Optimization of Image Quality and Organ Absorbed Dose for Pediatric Chest X-Ray Examination: In-House Developed Chest Phantom Study},
  journal = {Radiology Research and Practice},
  year = {2022},
  volume = {2022},
  pages = {1--10},
  doi = {10.1155/2022/3482458}
}
Kostenlos starten →