A Novel Spine Fixation System Made Entirely of Carbon-Fiber-Reinforced PEEK Composite: An In Vitro Mechanical Evaluation
Advances in Orthopedics · 09.06.2020 · 2020 · 1-7
DOI: 10.1155/2020/4796136 ↗Kurz gefasst
Fachartikel in der Zeitschrift Advances in Orthopedics, erschienen am 09.06.2020. Verfasst von Ofir Uri und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Hillel Yaffe Medical Center, Hadera, Israel. Die Arbeit wurde bisher 38-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CarboFix Orthopedics Ltd..
Abstract (Originalsprache)
Background
Semirigid spine fixation systems utilizing nonmetallic materials have emerged as a promising innovation to overcome the inherent disadvantages of metal instrumentation in spine surgery. This study tests the mechanical properties of a novel spine fixation system made entirely of carbon-fiber-reinforced PEEK (CFR-PEEK) composite material (CarboClear System, CarboFix Orthopedics Ltd., Israel).
Methods
An in vitro mechanical evaluation of the CFR-PEEK CarboClear system was conducted in compliance with the American Society for Testing and Materials (ASTM) F1717, F2193, and F543 standards.
Results
The mean bending yield load, bending ultimate load, and bending stiffness of the construct were 322 N, 363 N, and 45 N/mm, respectively. All tested samples completed 5 × 10 6 dynamic cycles successfully, with no evidence of fatigue failure at increasing load levels, up to 83% of ultimate bending load. The mean torsional stiffness was 1.0 Nm/deg and the mean screw axial pull-out strength was 2,037 N.
Conclusion
The CarboClear Pedicle Screw System has mechanical properties comparable to those of other commonly used titanium-made systems, with superior fatigue properties. The fatigue resistance, modulus of elasticity which is very similar to that of bone, radiolucency, and CT/MRI artifact-free feature of this spine fixation system made entirely of CFR-PEEK may offer advantages over traditional spine fixation systems made of metal alloys.
Autor:innen und Institutionen
- Department of Orthopedic Surgery, Hillel Yaffe Medical Center, Hadera, Israel.
ORCID-Profile: Ofir Uri, Eyal Behrbalk
Förderung
- CarboFix Orthopedics Ltd.
Zitieren
APA
Uri, O., Folman, Y., Laufer, G., & Behrbalk, E. (2020). A Novel Spine Fixation System Made Entirely of Carbon-Fiber-Reinforced PEEK Composite: An In Vitro Mechanical Evaluation. Advances in Orthopedics, 2020, 1-7. https://doi.org/10.1155/2020/4796136
Vancouver
Uri O, Folman Y, Laufer G, Behrbalk E. A Novel Spine Fixation System Made Entirely of Carbon-Fiber-Reinforced PEEK Composite: An In Vitro Mechanical Evaluation. Advances in Orthopedics. 2020;2020:1-7. doi:10.1155/2020/4796136
BibTeX
@article{Uri2020,
author = {Uri, Ofir and Folman, Yoram and Laufer, Gil and Behrbalk, Eyal},
title = {A Novel Spine Fixation System Made Entirely of Carbon-Fiber-Reinforced PEEK Composite: An In Vitro Mechanical Evaluation},
journal = {Advances in Orthopedics},
year = {2020},
volume = {2020},
pages = {1--7},
doi = {10.1155/2020/4796136}
}
EUVIMED