EUVIMEDEUVIMED entdecken
Open Access · CC BY 3.0

Evaluation of Motor Neuron Excitability by CMAP Scanning with Electric Modulated Current

Tiago Araújo, Rui Candeias, Neuza Nunes, Hugo Gamboa

Neuroscience Journal · 2015

DOI: 10.1155/2015/360648 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Neuroscience Journal, 2015. Verfasst von Tiago Araújo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Faculty of Sciences and Technology, New University of Lisbon, Lisbon, Portugal. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 8 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0). Gefördert von: Fundação para a Ciência e a Tecnologia, Universidade Nova de Lisboa.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Compound Muscle Action Potential (CMAP) scan is a noninvasive promissory technique for neurodegenerative pathologies diagnosis. In this work new CMAP scan protocols were implemented to study the influence of electrical pulse waveform on peripheral nerve excitability.

Methods

A total of 13 healthy subjects were tested. Stimulation was performed with an increasing intensities range from 4 to 30 mA. The procedure was repeated 4 times per subject, using a different single pulse stimulation waveform: monophasic square and triangular and quadratic and biphasic square.

Results

Different waveforms elicit different intensity-response amplitude curves. The square pulse needs less current to generate the same response amplitude regarding the other waves and this effect is gradually decreasing for the triangular, quadratic, and biphasic pulse, respectively.

Conclusion

The stimulation waveform has a direct influence on the stimulus-response slope and consequently on the motoneurons excitability. This can be a new prognostic parameter for neurodegenerative disorders.

Autor:innen und Institutionen

  • Department of Physics, Faculty of Sciences and Technology, New University of Lisbon, Lisbon, Portugal.
  • PLUX Wireless Biosignals S.A., Lisbon, Portugal.

Förderung

  • Fundação para a Ciência e a Tecnologia
  • Universidade Nova de Lisboa

Zitieren

APA

Araújo, T., Candeias, R., Nunes, N., & Gamboa, H. (2015). Evaluation of Motor Neuron Excitability by CMAP Scanning with Electric Modulated Current. Neuroscience Journal. https://doi.org/10.1155/2015/360648

Vancouver

Araújo T, Candeias R, Nunes N, Gamboa H. Evaluation of Motor Neuron Excitability by CMAP Scanning with Electric Modulated Current. Neuroscience Journal. 2015. doi:10.1155/2015/360648

BibTeX
@article{Arajo2015,
  author = {Araújo, Tiago and Candeias, Rui and Nunes, Neuza and Gamboa, Hugo},
  title = {Evaluation of Motor Neuron Excitability by CMAP Scanning with Electric Modulated Current},
  journal = {Neuroscience Journal},
  year = {2015},
  doi = {10.1155/2015/360648}
}
Kostenlos starten →