Fachartikel in der Zeitschrift Journal of Medical Engineering, erschienen am 24.03.2013. Verfasst von Fei Yang und 2 weiteren Autor:innen. Beteiligte Einrichtung: Washington University School of Medicine, Saint Louis, MO 63110, USA. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).
Abstract (Originalsprache)
Electrical impedance tomography (EIT) has the potential to provide a low cost and safe imaging modality for clinically monitoring patients being treated with mechanical ventilation. Variations in reconstruction algorithms at different clinical settings, however, make interpretation of regional ventilation across institutions difficult, presenting the need for a unified algorithm for thoracic EIT reconstruction. Development of such a consensual reconstruction algorithm necessitates a forward model capable of predicting surface impedance measurements as well as electric fields in the interior of the modeled thoracic volume. In this paper, we present an anatomically realistic forward solver for thoracic EIT that was built based on high resolution MR image data of a representative adult. Accuracy assessment of the developed forward solver in predicting surface impedance measurements by comparing the predicted and observed impedance measurements shows that the relative error is within the order of 5%, demonstrating the ability of the presented forward solver in generating high-fidelity surface thoracic impedance data for thoracic EIT algorithm development and evaluation.
Autor:innen und Institutionen
- Washington University School of Medicine, Saint Louis, MO 63110, USA.
- Division of Radiological Medical Physics, University of Kentucky, Lexington, KY 40536, USA.
- Institute of Engineering in Medicine, University of Minnesota, Minneapolis, MN 55455, USA.
Zitieren
APA
Yang, F., Zhang, J., & Patterson, R. (2013). Development of an Anatomically Realistic Forward Solver for Thoracic Electrical Impedance Tomography. Journal of Medical Engineering, 2013, 1-7. https://doi.org/10.1155/2013/983938
Vancouver
Yang F, Zhang J, Patterson R. Development of an Anatomically Realistic Forward Solver for Thoracic Electrical Impedance Tomography. Journal of Medical Engineering. 2013;2013:1-7. doi:10.1155/2013/983938
BibTeX
@article{Yang2013,
author = {Yang, Fei and Zhang, Jie and Patterson, Robert},
title = {Development of an Anatomically Realistic Forward Solver for Thoracic Electrical Impedance Tomography},
journal = {Journal of Medical Engineering},
year = {2013},
volume = {2013},
pages = {1--7},
doi = {10.1155/2013/983938}
}
EUVIMED