Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA
ISRN Obesity · 2013
DOI: 10.1155/2013/462394 ↗Kurz gefasst
Fachartikel in der Zeitschrift ISRN Obesity, 2013. Verfasst von Jean-Claude Pineau und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Human Evolution, French National Center for Scientific Research, UPR 2147-CNRS, 44 Amiral Mouchez Street, 75014 Paris, France. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).
Abstract (Originalsprache)
Objective
To examine measurement of body composition by ultrasound compared with a reference technique:dual energy X-ray absorptiometry (DXA). We evaluated the accuracy of a portable ultrasound-based device in estimating total body fat mass with those assessed by DXA in adult.
Methods
Body fat mass has been estimated using a portable ultrasound-based device in comparison with a contemporary reference DXA apparatus: the Hologic Discovery A. Anthropometric data has been assessed in order to maximize the output of the software associated with the ultrasound-based device. A cross-validation between ultrasound technique (US) and DXA was developed in this study. Total body fat mass estimated by ultrasound was compared with this DXA model in a sample of 83 women and 41 men.
Results
Ultrasound technique (US) of body fat (BF) was better correlated with DXA in both women ( r 2 = 0.97 , P < 0.01 ) and men ( r 2 = 0.92 , P < 0.01 ) with standard errors of estimates (SEE) being 2.1 kg and 2.2 kg, respectively.
Conclusion
The use of a portable device based on a US produced a very accurate BF estimate in relation to DXA reference technique. As DXA absorptiometry techniques are not interchangeable, the use of our ultrasound-based device needs to be recalibrated on a more contemporary DXA.
Autor:innen und Institutionen
- Department of Human Evolution, French National Center for Scientific Research, UPR 2147-CNRS, 44 Amiral Mouchez Street, 75014 Paris, France.
- Department of Public Health Sciences ,University of Modena and Reggio Emilia, Campi Road 287, 41100 Modena, Italy.
Zitieren
APA
Pineau, J. C., Lalys, L., Pellegrini, M., & Battistini, N. C. (2013). Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA. ISRN Obesity. https://doi.org/10.1155/2013/462394
Vancouver
Pineau JC, Lalys L, Pellegrini M, Battistini NC. Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA. ISRN Obesity. 2013. doi:10.1155/2013/462394
BibTeX
@article{Pineau2013,
author = {Pineau, Jean-Claude and Lalys, Loïc and Pellegrini, Massimo and Battistini, Nino Carlo},
title = {Body Fat Mass Assessment: A Comparison between an Ultrasound-Based Device and a Discovery A Model of DXA},
journal = {ISRN Obesity},
year = {2013},
doi = {10.1155/2013/462394}
}
EUVIMED