Comparing the In Vitro Stiffness of Straight-DCP, Wave-DCP, and LCP Bone Plates for Femoral Osteosynthesis
ISRN Orthopedics · 2013
DOI: 10.1155/2013/308753 ↗Kurz gefasst
Fachartikel in der Zeitschrift ISRN Orthopedics, 2013. Verfasst von José Ricardo Lenzi Mariolani und 1 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Biomaterials in Orthopedics, School of Medical Sciences, University of Campinas (UNICAMP), 126 Rua Tessália Vieira de Camargo, 13083-88…. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).
Abstract (Originalsprache)
The objective of this study was to compare the Locking Compression Plate (LCP) with the more cost-effective straight-dynamic compression plate (DCP) and wave-DCPs by testing in vitro the effects of plate stiffness on different types of diaphyseal femur fractures (A, B, and C, according to AO classification). The bending structural stiffness of each plate was obtained from four-point bending tests according to ASTM F382-99(2008). The plate systems were tested by applying compression/bending in different osteosynthesis simulation models using wooden rods to simulate the fractured bone fragments. Kruskal-Wallis test showed no significant difference in the bending structural stiffness between the three plate models. Rank-transformed two-way ANOVA showed significant influence of plate type, fracture type, and interaction plate versus fracture on the stiffness of the montages. The straight-DCP produced the most stable model for types B and C fractures, which makes its use advantageous for complex nonosteoporotic fractures that require minimizing focal mobility, whereas no difference was found for type A fracture. Our results indicated that DCPs, in straight or wave form, can provide adequate biomechanical properties for fixing diaphyseal femoral fractures in cases where more modern osteosynthesis systems are cost restrictive.
Autor:innen und Institutionen
- Laboratory of Biomaterials in Orthopedics, School of Medical Sciences, University of Campinas (UNICAMP), 126 Rua Tessália Vieira de Camargo, 13083-887 Campinas, SP, Brazil.
Zitieren
APA
Mariolani, J. R., & Belangero, W. D. (2013). Comparing the In Vitro Stiffness of Straight-DCP, Wave-DCP, and LCP Bone Plates for Femoral Osteosynthesis. ISRN Orthopedics. https://doi.org/10.1155/2013/308753
Vancouver
Mariolani JR, Belangero WD. Comparing the In Vitro Stiffness of Straight-DCP, Wave-DCP, and LCP Bone Plates for Femoral Osteosynthesis. ISRN Orthopedics. 2013. doi:10.1155/2013/308753
BibTeX
@article{Mariolani2013,
author = {Mariolani, José Ricardo Lenzi and Belangero, William Dias},
title = {Comparing the In Vitro Stiffness of Straight-DCP, Wave-DCP, and LCP Bone Plates for Femoral Osteosynthesis},
journal = {ISRN Orthopedics},
year = {2013},
doi = {10.1155/2013/308753}
}