Micromotion of Dental Implants: Basic Mechanical Considerations

Werner Winter, Daniel Klein, Matthias Karl

Journal of Medical Engineering · 20.11.2013 · 2013 · 1-9

DOI: 10.1155/2013/265412 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Medical Engineering, erschienen am 20.11.2013. Verfasst von Werner Winter und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, University of Erlangen-Nuremberg, Egerlandstraße 5, 91058 Erlangen, Germany. Die Arbeit wurde bisher 60-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Micromotion of dental implants may interfere with the process of osseointegration. Using three different types of virtual biomechanical models, varying contact types between implant and bone were simulated, and implant deformation, bone deformation, and stress at the implant-bone interface were recorded under an axial load of 200 N, which reflects a common biting force. Without friction between implant and bone, a symmetric loading situation of the bone with maximum loading and displacement at the apex of the implant was recorded. The addition of threads led to a decrease in loading and displacement at the apical part, but loading and displacement were also observed at the vertical walls of the implants. Introducing friction between implant and bone decreased global displacement. In a force fit situation, load transfer predominantly occurred in the cervical area of the implant. For freshly inserted implants, micromotion was constant along the vertical walls of the implant, whereas, for osseointegrated implants, the distribution of micromotion depended on the location. In the cervical aspect some minor micromotion in the range of 0.75 μ m could be found, while at the most apical part almost no relative displacement between implant and bone occurred.

Autor:innen und Institutionen

  • Department of Mechanical Engineering, University of Erlangen-Nuremberg, Egerlandstraße 5, 91058 Erlangen, Germany.
  • Department of Prosthodontics, University of Erlangen-Nuremberg, Glueckstraße 11, 91054 Erlangen, Germany.

Zitieren

APA

Winter, W., Klein, D., & Karl, M. (2013). Micromotion of Dental Implants: Basic Mechanical Considerations. Journal of Medical Engineering, 2013, 1-9. https://doi.org/10.1155/2013/265412

Vancouver

Winter W, Klein D, Karl M. Micromotion of Dental Implants: Basic Mechanical Considerations. Journal of Medical Engineering. 2013;2013:1-9. doi:10.1155/2013/265412

BibTeX
@article{Winter2013,
  author = {Winter, Werner and Klein, Daniel and Karl, Matthias},
  title = {Micromotion of Dental Implants: Basic Mechanical Considerations},
  journal = {Journal of Medical Engineering},
  year = {2013},
  volume = {2013},
  pages = {1--9},
  doi = {10.1155/2013/265412}
}
Kostenlos starten →