Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry

J Pan, A Saltos, D Smith, A Johnson, J Vossoughi

Journal of Medical Engineering · 2013

DOI: 10.1155/2013/165782 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Medical Engineering, 2013. Verfasst von J Pan und 4 weiteren Autor:innen. Beteiligte Einrichtung: Fischell Department of Bioengineering, University of Maryland, College Park, MD 20742, USA; Engineering and Scientific Research Associates, Olney, MD…. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0). Gefördert von: NHLBI NIH HHS, National Institutes of Health.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The airflow perturbation device (APD) has been developed as a portable, easy to use, and a rapid response instrument for measuring respiratory resistance in humans. However, the APD has limited data validating it against the established techniques. This study used a mechanical system to simulate the normal range of human breathing to validate the APD with the clinically accepted impulse oscillometry (IOS) technique. The validation system consisted of a sinusoidal flow generator with ten standardized resistance configurations that were shown to represent a total range of resistances from 0.12 to 0.95 kPa · L −1 · s (1.2–9.7 cm H 2 O · L −1 · s). Impulse oscillometry measurements and APD measurements of the mechanical system were recorded and compared at a constant airflow of 0.15 L · s −1 . Both the IOS and APD measurments were accurate in assessing nominal resistance. In addition, a strong linear relationship was observed between APD measurements and IOS measurements ( R 2 = 0.999). A second series of measurements was made on ten human volunteers with external resistors added in their respiratory flow paths. Once calibrated with the mechanical system, the APD gave respiratory resistance measurements within 5% of IOS measurements. Because of their comparability to IOS measurements, APD measurements are shown to be valid representations of respiratory resistance.

Autor:innen und Institutionen

  • Fischell Department of Bioengineering, University of Maryland, College Park, MD 20742, USA; Engineering and Scientific Research Associates, Olney, MD 20832, USA.
  • School of Medicine, University of Maryland, Baltimore, MD 21201, USA.
  • Engineering and Scientific Research Associates, Olney, MD 20832, USA.

Förderung

  • NHLBI NIH HHS
  • National Institutes of Health

Zitieren

APA

Pan, J., Saltos, A., Smith, D., Johnson, A., & Vossoughi, J. (2013). Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry. Journal of Medical Engineering. https://doi.org/10.1155/2013/165782

Vancouver

Pan J, Saltos A, Smith D, Johnson A, Vossoughi J. Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry. Journal of Medical Engineering. 2013. doi:10.1155/2013/165782

BibTeX
@article{Pan2013,
  author = {Pan, J and Saltos, A and Smith, D and Johnson, A and Vossoughi, J},
  title = {Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry},
  journal = {Journal of Medical Engineering},
  year = {2013},
  doi = {10.1155/2013/165782}
}
Kostenlos starten →