EUVIMEDEUVIMED entdecken
Vergleichsstudie

Exploratory Study of Radiation Exposure between Image Intensifier-Equipped C-Arm and a Flat Panel Detector During Volar Plate Fixation of the Distal Radius

Yuzo Honda, Eiji Kaida, Yoichi Miyazaki, Norihiko Sugita, Satoshi Mizukami, Kazuhiko Arima

The Journal of Hand Surgery (Asian-Pacific Volume) · 2026-08 · 31 (04) · 362-368

DOI: 10.1142/s2424835526500359 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Hand Surgery (Asian-Pacific Volume), online erschienen am 12.06.2026. Verfasst von Yuzo Honda und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Isahaya General Hospital, Isahaya, Nagasaki, Japan. Hauptthemen laut MEDLINE: Radiation Exposure, Radius Fractures, Wrist Fractures, Fracture Fixation, Internal. Die Arbeit nennt 26 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

To compare the amount of radiation exposure received by the surgeon during volar locking plate fixation for distal radius fractures using an image intensifier (II) and a flat panel detector (FPD).

Methods

We included 40 patients who underwent open reduction and internal fixation using a volar locking plate for isolated distal radius fractures. Surgeons wore thermoluminescent dosimeter badges at six anatomical sites: both thumbs, thyroid, at eye level (to estimate the eye lens dose), chest (under the protector) and abdomen (under the protector). The badges were replaced monthly, and equivalent radiation doses were measured. Fluoroscopy time was recorded from operative notes. Sixteen cases operated using an II-equipped C-arm between August and October 2020 were classified as Group II. Twenty-four patients treated between September and November 2024 using an FPD-equipped system comprised Group FPD.

Results

Operative time was similar between groups (57 vs. 56 minutes). Fluoroscopy time was 324 seconds in Group II and 239 seconds in Group FPD and air kerma per case was comparable (6.6 vs. 5.8 mGy). Equivalent dose per second was higher with FPD for both thumbs (right: 2.23 vs. 1.70 μSv/s; left: 9.57 vs. 2.07 μSv/s), with no significant differences in eye-lens or thyroid exposure.

Conclusions

The equivalent radiation dose per second of fluoroscopy was significantly higher in Group FPD for both thumbs compared to II systems. No significant differences were observed in radiation exposure to the lens or thyroid between groups. Level of Evidence: Level IV (Therapeutic)

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation ExposureRadius FracturesWrist FracturesFracture Fixation, InternalOccupational ExposureBone PlatesX-Ray Intensifying ScreensHumansFluoroscopyRadiation DosageMaleFemaleThermoluminescent DosimetryMiddle AgedPalmar PlateAged

Stichwörter der Autor:innen

Distal radius fracturesFracture fixationOrthopaedic fixation devicesOrthopaedic surgeonsRadiation exposure

Autor:innen und Institutionen

  • Department of Orthopedic Surgery, Isahaya General Hospital, Isahaya, Nagasaki, Japan.
  • Department of Orthopedic Surgery, Ainono Memorial Hospital, Ainocho Ko, Unzen, Nagasaki, Japan.
  • Department of Orthopedic Surgery, Ichinomiya Nishi Hospital, Ichinomiya, Aichi, Japan.
  • Department of Public Health, Nagasaki University Graduate School of Biomedical Sciences, Sakamoto, Nagasaki, Japan.

Zitieren

APA

Honda, Y., Kaida, E., Miyazaki, Y., Sugita, N., Mizukami, S., & Arima, K. (2026). Exploratory Study of Radiation Exposure between Image Intensifier-Equipped C-Arm and a Flat Panel Detector During Volar Plate Fixation of the Distal Radius. The Journal of Hand Surgery (Asian-Pacific Volume), 31(04), 362-368. https://doi.org/10.1142/s2424835526500359

Vancouver

Honda Y, Kaida E, Miyazaki Y, Sugita N, Mizukami S, Arima K. Exploratory Study of Radiation Exposure between Image Intensifier-Equipped C-Arm and a Flat Panel Detector During Volar Plate Fixation of the Distal Radius. The Journal of Hand Surgery (Asian-Pacific Volume). 2026;31(04):362-368. doi:10.1142/s2424835526500359

BibTeX
@article{Honda2026,
  author = {Honda, Yuzo and Kaida, Eiji and Miyazaki, Yoichi and Sugita, Norihiko and Mizukami, Satoshi and Arima, Kazuhiko},
  title = {Exploratory Study of Radiation Exposure between Image Intensifier-Equipped C-Arm and a Flat Panel Detector During Volar Plate Fixation of the Distal Radius},
  journal = {The Journal of Hand Surgery (Asian-Pacific Volume)},
  year = {2026},
  volume = {31},
  number = {04},
  pages = {362--368},
  doi = {10.1142/s2424835526500359}
}
Kostenlos starten →