Association of treponemal chemiluminescent immunoassay signal intensity with TPPA reactivity in reverse sequence syphilis screening
Microbiology Spectrum · 17.11.2026
DOI: 10.1128/spectrum.01501-26 ↗Kurz gefasst
Fachartikel in der Zeitschrift Microbiology Spectrum, online erschienen am 17.11.2026. Verfasst von Xin Liu und 9 weiteren Autor:innen. Beteiligte Einrichtung: Beijing YouAn Hospital, Capital Medical University, Beijing, China. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Reverse sequence screening for syphilis using treponemal chemiluminescent immunoassays (Tp-CIAs) is widely used in clinical laboratories; however, interpretation is typically based on binary cutoff values, and the clinical value of signal intensity remains underutilized. We conducted a retrospective diagnostic performance accuracy study including 1,827 individuals undergoing routine screening to evaluate the clinical utility of the Tp-CIA cutoff index (COI) as a continuous variable, using the TPPA as the reference standard. ROC analysis was performed to determine the optimal cutoff, and associations between COI values and TPPA reactivity were assessed. Tp-CIA demonstrated excellent discrimination of TPPA reactivity (area under the curve 0.992). COI values showed a strong monotonic association with TPPA reactivity (Spearman r s = 0.855), increasing from 2.5% at a COI
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Beijing YouAn Hospital, Capital Medical University, Beijing, China.
- School of Biomedical Sciences, Shandong First Medical University, Jinan, Shandong, China.
- Clinical Laboratory Center, Central Hospital of Enshi Tujia and Miao Autonomous Prefecture, Enshi, Hubei, China.
ORCID-Profile: Xin Liu, Shuping Tian, Jinli Lou, Hui Liu, Xia Feng, Yu Li, Shujing Huang, Yanhua Yu
Zitieren
APA
Liu, X., Tian, S., Lou, J., Liu, H., Feng, X., Li, Y., Jiang, F., Duan, Y., Huang, S., & Yu, Y. (2026). Association of treponemal chemiluminescent immunoassay signal intensity with TPPA reactivity in reverse sequence syphilis screening. Microbiology Spectrum. https://doi.org/10.1128/spectrum.01501-26
Vancouver
Liu X, Tian S, Lou J, Liu H, Feng X, Li Y, et al. Association of treponemal chemiluminescent immunoassay signal intensity with TPPA reactivity in reverse sequence syphilis screening. Microbiology Spectrum. 2026. doi:10.1128/spectrum.01501-26
BibTeX
@article{Liu2026,
author = {Liu, Xin and Tian, Shuping and Lou, Jinli and Liu, Hui and Feng, Xia and Li, Yu and Jiang, Feifei and Duan, Yuhan and Huang, Shujing and Yu, Yanhua},
title = {Association of treponemal chemiluminescent immunoassay signal intensity with TPPA reactivity in reverse sequence syphilis screening},
journal = {Microbiology Spectrum},
year = {2026},
doi = {10.1128/spectrum.01501-26}
}
EUVIMED