Comparative evaluation of the STANDARD M10 and Xpert C . difficile assays for detection of toxigenic Clostridioides difficile in stool specimens

Hyun-Woo Lee, Hui-Jin Yu, Heejung Kim, Sun Ae Yun, Eunsang Suh, Minhee Kang iD, Tae Yeul Kim iD, Hee Jae Huh iD, Nam Yong Lee iD

Journal of Clinical Microbiology · 2024

DOI: 10.1128/jcm.00524-24 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Clinical Microbiology, 2024. Verfasst von Hyun-Woo Lee und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea. Hauptthemen laut MEDLINE: Clostridioides difficile, Feces, Sensitivity and Specificity, Clostridium Infections. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: SD Biosensor.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study compared the performance of two commercial molecular assays, the STANDARD M10 Clostridioides difficile assay (M10) and the Xpert C. difficile assay (Xpert), for detecting toxigenic C. difficile in stool specimens. A total of 487 consecutive stool specimens submitted for routine C. difficile testing between June and November 2023 were included. Following routine testing using C. DIFF QUIK CHEK COMPLETE (QCC), M10 and Xpert were tested in parallel, alongside toxigenic culture (reference standard). Additionally, two-step algorithms, using QCC on the first step and either M10 or Xpert on the second step, were assessed. Both M10 and Xpert demonstrated a sensitivity and negative predictive value (NPV) of 100%. M10 exhibited significantly higher specificity and positive predictive value (PPV; 91.9% and 64.2%, respectively) than Xpert (90.3% and 59.8%, respectively). Both two-step algorithms showed a sensitivity and NPV of 98.4% and 99.8%, respectively. The specificity and PPV of the two-step algorithm using M10 (95.2% and 75.0%, respectively) were slightly higher than those of the one using Xpert (94.8% and 73.2%, respectively), without statistical significance. Receiver operating characteristic curve analysis, assessing the predictive ability of cycle threshold (Ct) values for the detection of free toxin, exhibited an area under the curve of 0.825 for M10 and 0.843 for Xpert. This indicates the utility of Ct values as predictors for the detection of free toxin in both assays. In conclusion, M10 proves to be an effective diagnostic tool with performance comparable to Xpert, whether utilized independently or as part of a two-step algorithm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Clostridioides difficileFecesSensitivity and SpecificityClostridium InfectionsMolecular Diagnostic TechniquesHumansAlgorithmsBacterial ToxinsPredictive Value of Tests

Stichwörter der Autor:innen

Clostridioides difficileM10Xpertcycle thresholdperformancetoxigenic culturetwo-step algorithm

Substanzen

Bacterial Toxins

Autor:innen und Institutionen

  • Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea.
  • Department of Laboratory Medicine, Seoul Medical Center, Seoul, South Korea.
  • Center for Clinical Medicine, Samsung Biomedical Research Institute, Samsung Medical Center, Seoul, South Korea.
  • Biomedical Engineering Research Center, Smart Healthcare Research Institute, Samsung Medical Center, Seoul, South Korea.
  • Department of Medical Device Management and Research, Samsung Advanced Institute for Health Sciences and Technology, Sungkyunkwan University, Seoul, South Korea.

ORCID-Profile: Minhee Kang, Tae Yeul Kim, Hee Jae Huh, Nam Yong Lee

Förderung

  • SD Biosensor

Zitieren

APA

Lee, H. W., Yu, H. J., Kim, H., Yun, S. A., Suh, E., Kang, M., Kim, T. Y., Huh, H. J., & Lee, N. Y. (2024). Comparative evaluation of the STANDARD M10 and Xpert C . difficile assays for detection of toxigenic Clostridioides difficile in stool specimens. Journal of Clinical Microbiology. https://doi.org/10.1128/jcm.00524-24

Vancouver

Lee HW, Yu HJ, Kim H, Yun SA, Suh E, Kang M, et al. Comparative evaluation of the STANDARD M10 and Xpert C . difficile assays for detection of toxigenic Clostridioides difficile in stool specimens. Journal of Clinical Microbiology. 2024. doi:10.1128/jcm.00524-24

BibTeX
@article{Lee2024,
  author = {Lee, Hyun-Woo and Yu, Hui-Jin and Kim, Heejung and Yun, Sun Ae and Suh, Eunsang and Kang, Minhee and Kim, Tae Yeul and Huh, Hee Jae and Lee, Nam Yong},
  title = {Comparative evaluation of the STANDARD M10 and Xpert C . difficile assays for detection of toxigenic Clostridioides difficile in stool specimens},
  journal = {Journal of Clinical Microbiology},
  year = {2024},
  doi = {10.1128/jcm.00524-24}
}