Performance of the Applied Biosystems HIV-1 Genotyping Kit with Integrase
Journal of Clinical Microbiology · 2024
DOI: 10.1128/jcm.00136-24 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Clinical Microbiology, 2024. Verfasst von Hannah P Moore und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pathology, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA. Hauptthemen laut MEDLINE: HIV-1, HIV Infections, Genotyping Techniques, Drug Resistance, Viral. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIAID NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
HIV genotyping is used to assess HIV susceptibility to antiretroviral drugs. The Applied Biosystems HIV-1 Genotyping Kit with Integrase (AB kit, Thermo Fisher Scientific) detects resistance-associated mutations (RAMs) in HIV protease (PR), reverse transcriptase (RT), and integrase (IN). We compared results from the AB kit with results obtained previously with the ViroSeq HIV-1 Genotyping System. DNA amplicons from the AB kit were also analyzed using next-generation sequencing (NGS). HIV RNA was extracted using the MagNA Pure 24 instrument (Roche Diagnostics; 96 plasma samples, HIV subtype B, viral load range: 530–737,741 copies/mL). FASTA files were generated from AB kit data using Exatype (Hyrax Biosciences). DNA amplicons from the AB kit were also analyzed by NGS using the Nextera XT kit (Illumina). Drug resistance was predicted using the Stanford HIV Drug Resistance Database. The mean genetic distance for sequences from ViroSeq and the AB kit was 0.02% for PR/RT and 0.04% for IN; 103 major RAMs were detected by both methods. Four additional major RAMs were detected by the AB kit only. These four major RAMs were also detected by NGS (detected in 18.1%–38.2% of NGS reads). NGS detected 27 major RAMs that were not detected with either of the Sanger sequencing-based kits. All major RAMs detected with ViroSeq were detected with the AB kit; additional RAMs were detected with the AB kit only. DNA amplicons from the AB kit can be used for NGS for more sensitive detection of RAMs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Pathology, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
- FHI 360, Durham, North Carolina, USA.
- Department of Medicine, Emory University School of Medicine, Atlanta, Georgia, USA.
- School of Social Work, Tulane Universtiy, New Orleans, Louisiana, USA.
- Department of Medicine, Harvard Medical School, Boston, Massachusetts, USA.
- Fenway Institute, Boston, Massachusetts, USA.
ORCID-Profile: Hannah P Moore, Susan H Eshleman
Förderung
- NIAID NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Moore, H. P., Palumbo, P. J., Notarte, K. I., Fogel, J. M., Cummings, V., Gamble, T., Del Rio, C., Batey, D. S., Mayer, K. H., Farley, J. E., Remien, R. H., Beyrer, C., Hudelson, S. E., Eshleman, S. H., & HPTN 078 study (2024). Performance of the Applied Biosystems HIV-1 Genotyping Kit with Integrase. Journal of Clinical Microbiology. https://doi.org/10.1128/jcm.00136-24
Vancouver
Moore HP, Palumbo PJ, Notarte KI, Fogel JM, Cummings V, Gamble T, et al. Performance of the Applied Biosystems HIV-1 Genotyping Kit with Integrase. Journal of Clinical Microbiology. 2024. doi:10.1128/jcm.00136-24
BibTeX
@article{Moore2024,
author = {Moore, Hannah P and Palumbo, Philip J and Notarte, Kin Israel and Fogel, Jessica M and Cummings, Vanessa and Gamble, Theresa and Del Rio, Carlos and Batey, D Scott and Mayer, Kenneth H and Farley, Jason E and Remien, Robert H and Beyrer, Chris and Hudelson, Sarah E and Eshleman, Susan H and HPTN 078 study, HPTN 078 study},
title = {Performance of the Applied Biosystems HIV-1 Genotyping Kit with Integrase},
journal = {Journal of Clinical Microbiology},
year = {2024},
doi = {10.1128/jcm.00136-24}
}
EUVIMED