Deployment of an electrocorticography system with a soft robotic actuator

Sukho Song iD, Florian Fallegger iD, Alix Trouillet, Kyungjin Kim iD, Stéphanie P Lacour iD

Science Robotics · 2023

DOI: 10.1126/scirobotics.add1002 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Science Robotics, 2023. Verfasst von Sukho Song und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory for Soft Bioelectronic Interfaces, Neuro-X Institute, ‌Ecole Polytechnique Fédérale de Lausanne (EPFL), 1202 Geneva, Switzerland. Hauptthemen laut MEDLINE: Robotics, Brain-Computer Interfaces. Die Arbeit wurde bisher 80-mal zitiert (Crossref) und nennt 67 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Electrocorticography (ECoG) is a minimally invasive approach frequently used clinically to map epileptogenic regions of the brain and facilitate lesion resection surgery and increasingly explored in brain-machine interface applications. Current devices display limitations that require trade-offs among cortical surface coverage, spatial electrode resolution, aesthetic, and risk consequences and often limit the use of the mapping technology to the operating room. In this work, we report on a scalable technique for the fabrication of large-area soft robotic electrode arrays and their deployment on the cortex through a square-centimeter burr hole using a pressure-driven actuation mechanism called eversion. The deployable system consists of up to six prefolded soft legs, and it is placed subdurally on the cortex using an aqueous pressurized solution and secured to the pedestal on the rim of the small craniotomy. Each leg contains soft, microfabricated electrodes and strain sensors for real-time deployment monitoring. In a proof-of-concept acute surgery, a soft robotic electrode array was successfully deployed on the cortex of a minipig to record sensory cortical activity. This soft robotic neurotechnology opens promising avenues for minimally invasive cortical surgery and applications related to neurological disorders such as motor and sensory deficits.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RoboticsBrain-Computer InterfacesAnimalsSwineElectrocorticographySwine, MiniatureBrain

Autor:innen und Institutionen

  • Laboratory for Soft Bioelectronic Interfaces, Neuro-X Institute, ‌Ecole Polytechnique Fédérale de Lausanne (EPFL), 1202 Geneva, Switzerland.
  • Laboratory of Sustainability Robotics, Swiss Federal Laboratories for Materials Science and Technology (Empa), 8600 Dübendorf, Switzerland.
  • Department of Mechanical Engineering, University of Connecticut, Storrs, CT 06269, USA.

ORCID-Profile: Sukho Song, Florian Fallegger, Kyungjin Kim, Stéphanie P Lacour

Zitieren

APA

Song, S., Fallegger, F., Trouillet, A., Kim, K., & Lacour, S. P. (2023). Deployment of an electrocorticography system with a soft robotic actuator. Science Robotics. https://doi.org/10.1126/scirobotics.add1002

Vancouver

Song S, Fallegger F, Trouillet A, Kim K, Lacour SP. Deployment of an electrocorticography system with a soft robotic actuator. Science Robotics. 2023. doi:10.1126/scirobotics.add1002