EUVIMEDEUVIMED entdecken
Crossref

Validity and Reliability of the 3-Dimensional Printed PhysioCode Handheld Dynamometer for Knee Muscle Strength Assessment

George Schayer Sabino iD, Diego Henrique Antunes Nascimento iD, Fernanda Oliveira Madaleno iD, Claysson Bruno Santos Vimieiro iD, Pollyana Helena Vieira Costa iD, Guilherme Augusto Santos Araujo iD, Matheus Hissa Lourenço Ferreira iD, Renan Alves Resende iD

Journal of Sport Rehabilitation · 2026 · 1-7

DOI: 10.1123/jsr.2025-0408 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Sport Rehabilitation, 2026. Verfasst von George Schayer Sabino und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physical Therapy, Graduate Program in Rehabilitation Sciences, School of Physical Education, Physical Therapy and Occupational Therapy,…. Die Arbeit nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Context

Handheld dynamometry is common for isometric strength testing, but the high cost and import barriers of commercial devices (eg, MicroFET II) limit access. Three-dimensional (3D) printing enables low-cost, customizable instruments. We developed the 3D-printed PhysioCode and examined its measurement properties for knee strength.

Design

Methods and Technical Report evaluating concurrent validity and test–retest reliability versus a reference handheld dynamometer.

Methods

Forty-nine healthy adults (23.2 [3.9] y; body mass index 21.9 [2.8] kg/m 2 ) performed 3 isometric trials of dominant-limb knee extension and flexion with both devices in randomized order, assessed by 2 trained novice raters under blinded analysis. Concurrent validity was evaluated using Pearson correlation coefficients ( r ) and root mean square error, and agreement between devices was assessed using Bland–Altman analyses. Reliability was assessed using intraclass correlation coefficients (ICC[2,1] and ICC[2, k ]), along with the standard error of measurement and minimal detectable change at 95% confidence. We compared single and repeated test strategies.

Results

All mean values differed by ≤21 N between devices across knee extensors and flexors. Concurrent validity was strong–very strong ( r = .88–.95; P < .001), with lower root mean square error for averaged repetitions. Bland–Altman showed small, direction-specific bias: PhysioCode slightly overestimated extensor force values (−17 to 21 N) and underestimated flexor force values (−12 N). Reliability was excellent: PhysioCode ICC(2,1) = .93 to .95 and ICC(2, k ) = .98 to .98; MicroFET II ICC(2,1) = .89 to .94 and ICC(2, k ) = .96 to .98. Averaging 2 to 3 trials reduced standard error of measurement and minimal detectable change at 95% confidence relative to single trials or maxima.

Conclusions

The 3D-printed PhysioCode showed adequate concurrent validity and excellent test–retest reliability for knee extensor/flexor strength, with performance comparable to the MicroFET II in this sample of young, healthy adults. Averaging 2 or 3 repetitions improved precision. These findings suggest that PhysioCode may represent an option for strength assessment, warranting further evaluation in target clinical populations.

Schlagwörter

Stichwörter der Autor:innen

3Dknee jointprintingreproducibility of results

Autor:innen und Institutionen

  • Department of Physical Therapy, Graduate Program in Rehabilitation Sciences, School of Physical Education, Physical Therapy and Occupational Therapy, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG, Braz…
  • Faculdade Ciências Médicas de Minas Gerais (FCMMG), Belo Horizonte, MG, Brazil.
  • Bioengineering Laboratory, Graduate Program in Mechanical Engineering, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG, Brazil.
  • Faculdade de Sete Lagoas (FACSETE), Sete Lagoas, MG, Brazil.
  • Department of Sport Sciences, School of Physical Education, Physical Therapy and Occupational Therapy, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG, Brazil.

ORCID-Profile: George Schayer Sabino, Diego Henrique Antunes Nascimento, Fernanda Oliveira Madaleno, Claysson Bruno Santos Vimieiro, Pollyana Helena Vieira Costa, Guilherme Augusto Santos Araujo, Matheus Hissa Lourenço Ferreira, Renan Alves Resende

Zitieren

APA

Sabino, G. S., Nascimento, D. H. A., Madaleno, F. O., Vimieiro, C. B. S., Costa, P. H. V., Araujo, G. A. S., Ferreira, M. H. L., & Resende, R. A. (2026). Validity and Reliability of the 3-Dimensional Printed PhysioCode Handheld Dynamometer for Knee Muscle Strength Assessment. Journal of Sport Rehabilitation, 1-7. https://doi.org/10.1123/jsr.2025-0408

Vancouver

Sabino GS, Nascimento DHA, Madaleno FO, Vimieiro CBS, Costa PHV, Araujo GAS, et al. Validity and Reliability of the 3-Dimensional Printed PhysioCode Handheld Dynamometer for Knee Muscle Strength Assessment. Journal of Sport Rehabilitation. 2026:1-7. doi:10.1123/jsr.2025-0408

BibTeX
@article{Sabino2026,
  author = {Sabino, George Schayer and Nascimento, Diego Henrique Antunes and Madaleno, Fernanda Oliveira and Vimieiro, Claysson Bruno Santos and Costa, Pollyana Helena Vieira and Araujo, Guilherme Augusto Santos and Ferreira, Matheus Hissa Lourenço and Resende, Renan Alves},
  title = {Validity and Reliability of the 3-Dimensional Printed PhysioCode Handheld Dynamometer for Knee Muscle Strength Assessment},
  journal = {Journal of Sport Rehabilitation},
  year = {2026},
  pages = {1--7},
  doi = {10.1123/jsr.2025-0408}
}
Kostenlos starten →