Julianna R Voelker iD, Devon Kulinski iD, Douglas S Brungart iD

The Journal of the Acoustical Society of America · 01.04.2026 · 159 (4) · 3103-3113

DOI: 10.1121/10.0042977 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The Journal of the Acoustical Society of America, online erschienen am 08.04.2026. Verfasst von Julianna R Voelker und 2 weiteren Autor:innen. Beteiligte Einrichtung: Audiology and Speech Center, Walter Reed National Military Medical Center, Bethesda, Maryland 20889, USA. Hauptthemen laut MEDLINE: Neck, Ear Protective Devices, Acoustics. Die Arbeit nennt 18 Literaturangaben. Gefördert von: Military Operational Medicine Research Program, Telemedicine and Advanced Technology Research Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Many commercially available hearing protector fit testing (HPFT) systems are able to measure personal attenuation rating (PAR) values for earplugs, but only a few are capable of measuring PAR with over-the-ear protectors and even fewer are able to estimate PAR values with double hearing protection. One approach that could potentially be used with any type of hearing protection is to modify the real-ear attenuation at threshold method to use an occluded threshold that is measured with circumaural headphones worn around the neck rather than over the ears. However, this method would only be valid if the neck-to-ear transfer function of the headphones is stable across a variety of headphone fittings. This study evaluated the validity of a neck-based HPFT approach by (1) measuring the transfer function from headphones worn around the neck to the ear across a variety of headphone positions and head and ear geometries; and (2) using an acoustic test fixture to compare PAR values measured with a neck-based sound source to those obtained with more conventional HPFT approaches. The results indicate that the neck-based HPFT system has a relatively consistent neck-to-ear transfer function and produced PAR values comparable to those obtained with other HPFT systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

NeckEar Protective DevicesAcousticsHumansReproducibility of ResultsEquipment DesignAuditory ThresholdAcoustic StimulationNoise

Autor:innen und Institutionen

  • Audiology and Speech Center, Walter Reed National Military Medical Center, Bethesda, Maryland 20889, USA.
  • Henry M. Jackson Foundation, Bethesda, Maryland 20817, USA.
  • Po'okela Government Solutions LLC, Frederick, Maryland 21704, USA.

ORCID-Profile: Julianna R Voelker, Devon Kulinski, Douglas S Brungart

Förderung

  • Military Operational Medicine Research Program
  • Telemedicine and Advanced Technology Research Center

Zitieren

APA

Voelker, J. R., Kulinski, D., & Brungart, D. S. (2026). Electroacoustic validation of a neck-based field attenuation estimation system. The Journal of the Acoustical Society of America, 159(4), 3103-3113. https://doi.org/10.1121/10.0042977

Vancouver

Voelker JR, Kulinski D, Brungart DS. Electroacoustic validation of a neck-based field attenuation estimation system. The Journal of the Acoustical Society of America. 2026;159(4):3103-3113. doi:10.1121/10.0042977

BibTeX
@article{Voelker2026,
  author = {Voelker, Julianna R and Kulinski, Devon and Brungart, Douglas S},
  title = {Electroacoustic validation of a neck-based field attenuation estimation system},
  journal = {The Journal of the Acoustical Society of America},
  year = {2026},
  volume = {159},
  number = {4},
  pages = {3103--3113},
  doi = {10.1121/10.0042977}
}
Kostenlos starten →