Modeling peristaltic flow dynamics with ultrasound localization microscopy: An in vitro study
The Journal of the Acoustical Society of America · 01.10.2025 · 158 (4_Supplement) · A41-A41
DOI: 10.1121/10.0040060 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of the Acoustical Society of America, erschienen am 01.10.2025. Verfasst von Luca Giaccone und 2 weiteren Autor:innen.
Abstract (Originalsprache)
By tracking microbubbles (MBs) in the bloodstream, ultrasound localization microscopy (ULM) reconstructs high-resolution velocity maps of vascular networks. While ULM offers a promising alternative to conventional Doppler techniques, its raw output is a large, noisy set of MB trajectories, making it challenging to extract reliable and interpretable velocity estimates. Considering the flow dynamics has been shown to improve interpretability and support the extraction of clinically relevant biomarkers, such as the Pulsatility Index (PI). However, previous studies have been limited to analyses at the level of vessel mean velocity. In this work, we study the peristaltic flow dynamics throughout the entire radial velocity profile in a cylindrical channel. We evaluate six models on an in vitro dataset, including both steady (Poiseuille) and unsteady (Womersley) solutions, under no-slip and slip boundary conditions. Our results show that accounting for pulsatility and boundary slip significantly improves the fit to experimental data. As a consequence, this approach also provides a more robust PI estimation, computed from the total channel flow rather than the centerline velocity alone. These findings highlight that physics-informed modeling can enhance the extraction of meaningful flow characteristics from ULM data, advancing the development of more reliable quantitative biomarkers and improving measurements interpretation.
Zitieren
APA
Giaccone, L., Tuccio, G., & Demi, L. (2025). Modeling peristaltic flow dynamics with ultrasound localization microscopy: An in vitro study. The Journal of the Acoustical Society of America, 158(4_Supplement), A41-A41. https://doi.org/10.1121/10.0040060
Vancouver
Giaccone L, Tuccio G, Demi L. Modeling peristaltic flow dynamics with ultrasound localization microscopy: An in vitro study. The Journal of the Acoustical Society of America. 2025;158(4_Supplement):A41-A41. doi:10.1121/10.0040060
BibTeX
@article{Giaccone2025,
author = {Giaccone, Luca and Tuccio, Giulia and Demi, Libertario},
title = {Modeling peristaltic flow dynamics with ultrasound localization microscopy: An in vitro study},
journal = {The Journal of the Acoustical Society of America},
year = {2025},
volume = {158},
number = {4_Supplement},
pages = {A41--A41},
doi = {10.1121/10.0040060}
}