Smartphone tristimulus colorimetry for skin-tone analysis at common pulse oximetry anatomical sites

Joshua A Burrow iD, Rutendo Jakachira iD, Gannon Lemaster, Kimani C Toussaint

Biophotonics Discovery · 19.05.2025 · 2 (03) · 1-13

DOI: 10.1117/1.bios.2.3.032504 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biophotonics Discovery, erschienen am 19.05.2025. Verfasst von Joshua A Burrow und 3 weiteren Autor:innen. Beteiligte Einrichtung: Morgan State University, Electrical & Computer Engineering Department, Baltimore, Maryland, United States. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

SignificanceSmartphones hold great potential in point-of-care settings due to their accessibility and computational capabilities. This is critical as clinicians increasingly seek to quantify skin-tone, a characteristic that has been shown to impact the accuracy of pulse oximetry readings, particularly for dark skin tones, and, hence, disproportionately affect patient outcomes.AimWe present a smartphone-based imaging technique for determining individual typology angle (ITA) and compare these results to those obtained using an industry-standard tristimulus colorimeter, particularly for the finger, a common site for pulse oximetry measurements.ApproachWe employ a smartphone-based imaging method to extract ITA values from four volunteers across diverse skin-tones. We provide recommendations for minimizing errors caused by ambient light scattering, which can affect skin-tone readings.ResultsThe smartphone-based ITA (SITA) measurements with camera flash disabled and minimal ambient lighting correlate well with an industry-standard colorimeter without the need for auxiliary adapters and complex calibration. The method presented enables wide-field ITA mapping for skin-tone quantification that is accessible to clinicians.ConclusionsOur findings demonstrate that smartphone-based imaging provides an effective alternative for assessing skin-tone in clinical settings. The reduced complexity of the approach presented makes it highly accessible to the clinical community and others interested in carrying out pulse oximetry across a diversity of skin-tones in a manner that standardizes skin-tone assessment.

Schlagwörter

Stichwörter der Autor:innen

colorimetryequitable healthindividual typology anglepulse oximetryskin-tonesmartphone

Autor:innen und Institutionen

  • Morgan State University, Electrical & Computer Engineering Department, Baltimore, Maryland, United States.
  • Brown University, School of Engineering, PROBE Lab, Providence, Rhode Island, United States.
  • Brown University, Physics Department, Providence, Rhode Island, United States.
  • Brown University, Center for Digital Health, Providence, Rhode Island, United States.

ORCID-Profile: Joshua A Burrow, Rutendo Jakachira

Zitieren

APA

Burrow, J. A., Jakachira, R., Lemaster, G., & Toussaint, K. C. (2025). Smartphone tristimulus colorimetry for skin-tone analysis at common pulse oximetry anatomical sites. Biophotonics Discovery, 2(03), 1-13. https://doi.org/10.1117/1.bios.2.3.032504

Vancouver

Burrow JA, Jakachira R, Lemaster G, Toussaint KC. Smartphone tristimulus colorimetry for skin-tone analysis at common pulse oximetry anatomical sites. Biophotonics Discovery. 2025;2(03):1-13. doi:10.1117/1.bios.2.3.032504

BibTeX
@article{Burrow2025,
  author = {Burrow, Joshua A and Jakachira, Rutendo and Lemaster, Gannon and Toussaint, Kimani C},
  title = {Smartphone tristimulus colorimetry for skin-tone analysis at common pulse oximetry anatomical sites},
  journal = {Biophotonics Discovery},
  year = {2025},
  volume = {2},
  number = {03},
  pages = {1--13},
  doi = {10.1117/1.bios.2.3.032504}
}
Kostenlos starten →