Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen
Journal of Vacuum Science & Technology B · 01.12.2026 · 44 (6)
DOI: 10.1116/6.0005912 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Vacuum Science & Technology B, online erschienen am 19.11.2026. Verfasst von J. Zakrisson und 3 weiteren Autor:innen. Die Arbeit nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Partnership on Metrology, Vetenskapsrådet, VINNOVA.
Abstract (Originalsprache)
Two Fabry–Pérot (FP) refractometers for the realization of the pascal, a novel single-cavity system and a well-established dual-cavity system, are independently characterized and compared. Both systems operate with N2 at pressures up to 100 kPa and are evaluated using molecular data that provide traceability to helium. The two systems yield comparable expanded uncertainties (k = 2), given by [(1.6 mPa)2 + (13 × 10−6p)2 + (37 × 10−12p2/Pa)2]1/2 for the single-cavity system and [(1.8 mPa)2 + (13 × 10−6p)2 + (37 × 10−12p2/Pa)2]1/2 for the dual-cavity system, demonstrating that the novel single-cavity architecture can achieve a pressure realization performance essentially equivalent to that of the established dual-cavity system. For both systems, the uncertainty is dominated by the uncertainties associated with the molar refractivity and second density virial coefficient of N2. A direct comparison of the two systems using the diffusion-isolated pressure equilibration methodology shows excellent agreement, with systematic linear and quadratic response differences of 0.17 × 10−6p and 9.2 × 10−12p2/Pa, respectively. Over an 18-day measurement campaign, the differential assessments exhibit a repeatability of 0.6 × 10−6 (k = 2), with virtually no detectable drift. These results demonstrate consistent performance between two independently realized refractometers based on slightly different cavity architectures, supporting the robustness of FP-based refractometry for high-accuracy pressure realization. They also highlight the need for improved determinations of molecular parameters of N2 that are traceable to helium, particularly its molar refractivity and second density virial coefficient.
Förderung
- European Partnership on Metrology
- Vetenskapsrådet
- VINNOVA
Zitieren
APA
Zakrisson, J., Silander, I., Axner, O., & Zelan, M. (2026). Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen. Journal of Vacuum Science & Technology B, 44(6). https://doi.org/10.1116/6.0005912
Vancouver
Zakrisson J, Silander I, Axner O, Zelan M. Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen. Journal of Vacuum Science & Technology B. 2026;44(6). doi:10.1116/6.0005912
BibTeX
@article{Zakrisson2026,
author = {Zakrisson, J. and Silander, I. and Axner, O. and Zelan, M.},
title = {Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen},
journal = {Journal of Vacuum Science & Technology B},
year = {2026},
volume = {44},
number = {6},
doi = {10.1116/6.0005912}
}
EUVIMED