Assessing dead time effects when attempting isotope ratio quantification by time-of-flight secondary ion mass spectrometry
Biointerphases · 2023
DOI: 10.1116/6.0002954 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biointerphases, 2023. Verfasst von Laura C Baqué und 3 weiteren Autor:innen. Beteiligte Einrichtung: Nanoscience and Nanotechnology Institute (CNEA-CONICET), Department of Materials Characterization, Bariloche Atomic Center, Av. Bustillo 9500, S. C. …. Hauptthemen laut MEDLINE: Spectrometry, Mass, Secondary Ion, Isotopes. Die Arbeit nennt 17 Literaturangaben. Gefördert von: Agencia Nacional de Promoción Científica y Tecnológica.
Abstract (Originalsprache)
Time-of-flight secondary ion mass spectrometry (TOF-SIMS) is a quasi-non-destructive technique capable of analyzing the outer monolayers of a solid sample and detecting all elements of the periodic table and their isotopes. Its ability to analyze the outer monolayers resides in sputtering the sample surface with a low-dose primary ion gun, which, in turn, imposes the use of a detector capable of counting a single ion at a time. Consequently, the detector saturates when more than one ion arrives at the same time hindering the use of TOF-SIMS for quantification purposes such as isotope ratio estimation. Even though a simple Poisson-based correction is usually implemented in TOF-SIMS acquisition software to compensate the detector saturation effects, this correction is only valid up to a certain extent and can be unnoticed by the inexperienced user. This tutorial describes a methodology based on different practices reported in the literature for dealing with the detector saturation effects and assessing the validity limits of Poisson-based correction when attempting to use TOF-SIMS data for quantification purposes. As a practical example, a dried lithium hydroxide solution was analyzed by TOF-SIMS with the aim of estimating the 6Li/7Li isotope ratio. The approach presented here can be used by new TOF-SIMS users on their own data for understanding the effects of detector saturation, determine the validity limits of Poisson-based correction, and take into account important considerations when treating the data for quantification purposes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Nanoscience and Nanotechnology Institute (CNEA-CONICET), Department of Materials Characterization, Bariloche Atomic Center, Av. Bustillo 9500, S. C. de Bariloche, Río Negro R8402AGP, Argentina.
- Nanoscience and Nanotechnology Institute (CNEA-CONICET), Department of Condensed Matter Physics, Constituyentes Atomic Center, Av. General Paz 1499, San Martín, Buenos Aires B1650KNA, Argentina.
- Argentine Neutron Beam Laboratory (LAHN), CNEA, Av. General Paz 1499, San Martín, Buenos Aires B1650KNA, Argentina.
ORCID-Profile: Laura C Baqué, Federico M Cabello, Federico A Viva, Horacio R Corti
Förderung
- Agencia Nacional de Promoción Científica y Tecnológica
Zitieren
APA
Baqué, L. C., Cabello, F. M., Viva, F. A., & Corti, H. R. (2023). Assessing dead time effects when attempting isotope ratio quantification by time-of-flight secondary ion mass spectrometry. Biointerphases. https://doi.org/10.1116/6.0002954
Vancouver
Baqué LC, Cabello FM, Viva FA, Corti HR. Assessing dead time effects when attempting isotope ratio quantification by time-of-flight secondary ion mass spectrometry. Biointerphases. 2023. doi:10.1116/6.0002954
BibTeX
@article{Baqu2023,
author = {Baqué, Laura C and Cabello, Federico M and Viva, Federico A and Corti, Horacio R},
title = {Assessing dead time effects when attempting isotope ratio quantification by time-of-flight secondary ion mass spectrometry},
journal = {Biointerphases},
year = {2023},
doi = {10.1116/6.0002954}
}
EUVIMED