Comparative efficacy of conventional and handheld ultrasound devices in detecting bone fractures: an in vitro study

Mohammed J Alsaadi, Abdullah Khalid Alhejji, Abdulaziz Mohammed Alkhamis, Sultan Turki Alshehri, Ziyad Abdulaziz Almutiri, Salman Mohammad Alshulail, Malak Almutairi, Norah Ahmad Althrwi, Abdulrahman M Alfuraih

Medical Ultrasonography · 2026

DOI: 10.11152/mu-4559 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Medical Ultrasonography, 2026. Verfasst von Mohammed J Alsaadi und 8 weiteren Autor:innen. Beteiligte Einrichtung: Radiology and Medical Imaging Department, College of Applied Medical Sciences, Prince Sattam Bin Abdulaziz University. m.alsaadi@psau.edu.sa. Hauptthemen laut MEDLINE: Fractures, Bone. Die Arbeit wurde bisher 1-mal zitiert (Crossref).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

Ultrasound (US) is increasingly used for fracture diagnosis, yet the comparative performance of handheld and conventional US, particularly by novice users, remains unclear. This study evaluated their diagnostic accuracy using in vitro bone models.Material and methods: Five blinded junior sonographers examined seven bone samples (comprising four fractured, comminuted, transverse, obliquely displaced, and linear, as well as three intact samples) using both types of devices. Diagnostic sensitivity, specificity, and accuracy were calculated. McNemar’s test evaluated differences in performance, Cohen’s Kappa measured inter-device agreement, and McNemar’s test along with mixed-effects logistic regression analysed the impact of fracture types. Inter-sonographer reliability was assessed using the intraclass correlation coefficient (ICC).

Results

Both devices showed 85% sensitivity (95% CI: 61.1–96%). Specificity was low (conventional: 20.0%; handheld: 33.3%). Accuracy was moderate (conventional: 57.1%; handheld: 62.9%). No significant difference was found (p = .21), and inter-device agreement was poor (Kappa = 0.05±0.20). Comminuted fractures were significantly more detectable than intact bones (OR=2.25, p=0.03). Inter-operator reliability was low (ICC: 0.28–0.35).

Conclusion

Handheld US showed slightly higher specificity and accuracy, but both systems demonstrated limited agreement and variable novice performance. These findings highlight the need for improved training and optimisation when using US for fracture detection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fractures, BoneUltrasonographySensitivity and SpecificityReproducibility of ResultsHumansEquipment DesignIn Vitro Techniques

Autor:innen und Institutionen

  • Radiology and Medical Imaging Department, College of Applied Medical Sciences, Prince Sattam Bin Abdulaziz University. m.alsaadi@psau.edu.sa.
  • Radiology and Medical Imaging Department, College of Applied Medical Sciences, Prince Sattam Bin Abdulaziz University.
  • Alkarj Armed Forces Hospital, Ministry of Defense.
  • Prince Salman Bin Muhammad Hospital in Dalam Governorate, Al-Kharj.

Zitieren

APA

Alsaadi, M. J., Alhejji, A. K., Alkhamis, A. M., Alshehri, S. T., Almutiri, Z. A., Alshulail, S. M., Almutairi, M., Althrwi, N. A., & Alfuraih, A. M. (2026). Comparative efficacy of conventional and handheld ultrasound devices in detecting bone fractures: an in vitro study. Medical Ultrasonography. https://doi.org/10.11152/mu-4559

Vancouver

Alsaadi MJ, Alhejji AK, Alkhamis AM, Alshehri ST, Almutiri ZA, Alshulail SM, et al. Comparative efficacy of conventional and handheld ultrasound devices in detecting bone fractures: an in vitro study. Medical Ultrasonography. 2026. doi:10.11152/mu-4559

BibTeX
@article{Alsaadi2026,
  author = {Alsaadi, Mohammed J and Alhejji, Abdullah Khalid and Alkhamis, Abdulaziz Mohammed and Alshehri, Sultan Turki and Almutiri, Ziyad Abdulaziz and Alshulail, Salman Mohammad and Almutairi, Malak and Althrwi, Norah Ahmad and Alfuraih, Abdulrahman M},
  title = {Comparative efficacy of conventional and handheld ultrasound devices in detecting bone fractures: an in vitro study},
  journal = {Medical Ultrasonography},
  year = {2026},
  doi = {10.11152/mu-4559}
}
Kostenlos starten →