Computational Modeling of Orthopedic Screw Pullout Performance According to ASTM F543: A Validation Approach Based on ASME V&V40 Standard and Food and Drug Administration Guidance
Journal of Verification, Validation and Uncertainty Quantification · 01.09.2025 · 10 (3)
DOI: 10.1115/1.4071436 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Verification, Validation and Uncertainty Quantification, online erschienen am 06.04.2026. Verfasst von David Benoit und 6 weiteren Autor:innen. Die Arbeit nennt 36 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council of Canada.
Abstract (Originalsprache)
Computational modeling and simulation (CM&S) is increasingly used to complement or replace traditional bench testing in the design, evaluation, and regulatory assessment of orthopedic devices. For orthopedic bone screws, pullout performance is commonly assessed through ASTM F543 testing, but this process can be costly and time-consuming. Finite element (FE) models, when subjected to rigorous verification, validation, and uncertainty quantification, can serve as surrogates to reduce reliance on experimental testing while supporting regulatory submissions. This study presents an end-to-end methodology for establishing the credibility of a device-agnostic FE model simulating screw pullout in accordance with ASTM F543. The model replicates the mechanical interaction between a bone screw and synthetic bone foam, with the quantity of interest being the maximum pullout force. Its context of use (CoU) is defined as a full surrogate for physical testing, and its associated use risk was determined to be medium-high, based on its role in device safety and performance assessment. Credibility activities were defined using the ASME V&V40 framework and the 2023 Food and Drug Administration (FDA) CM&S credibility guidance. Twenty-five credibility factors were reviewed, and high ratings were achieved for 19, reflecting a high degree of rigor across verification, calibration, validation, and applicability evidence. This work illustrates how a risk-informed credibility framework can be applied to establish a validated computational model of a standardized test in the medical device field, supporting its use as a surrogate for bench testing in both development and regulatory contexts.
Förderung
- Natural Sciences and Engineering Research Council of Canada
Zitieren
APA
Benoit, D., Navéaux, F. L., Degueldre, L., Beauséjour, M., Petit, Y., Wagnac, É., & Clin, J. (2025). Computational Modeling of Orthopedic Screw Pullout Performance According to ASTM F543: A Validation Approach Based on ASME V&V40 Standard and Food and Drug Administration Guidance. Journal of Verification, Validation and Uncertainty Quantification, 10(3). https://doi.org/10.1115/1.4071436
Vancouver
Benoit D, Navéaux FL, Degueldre L, Beauséjour M, Petit Y, Wagnac É, et al. Computational Modeling of Orthopedic Screw Pullout Performance According to ASTM F543: A Validation Approach Based on ASME V&V40 Standard and Food and Drug Administration Guidance. Journal of Verification, Validation and Uncertainty Quantification. 2025;10(3). doi:10.1115/1.4071436
BibTeX
@article{Benoit2025,
author = {Benoit, David and Navéaux, Franck Le and Degueldre, Loïc and Beauséjour, Marie-Hélène and Petit, Yvan and Wagnac, Éric and Clin, Julien},
title = {Computational Modeling of Orthopedic Screw Pullout Performance According to ASTM F543: A Validation Approach Based on ASME V&V40 Standard and Food and Drug Administration Guidance},
journal = {Journal of Verification, Validation and Uncertainty Quantification},
year = {2025},
volume = {10},
number = {3},
doi = {10.1115/1.4071436}
}