Biomechanics of Manual-Bent Versus Patient-Specific Mandibular Implants
Journal of Biomechanical Engineering · 2025
DOI: 10.1115/1.4069665 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Biomechanical Engineering, 2025. Verfasst von Shengping Zhong und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging & Pathology, Faculty of Medicine, KU Leuven, Kapucijnenvoer 7, Box 7003, Leuven 3000, Belgium;Oral and Maxillofacial Surgery, U…. Hauptthemen laut MEDLINE: Mandible, Mechanical Phenomena. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben.
Abstract (Originalsprache)
The patient-specific mandibular reconstruction plate (PSMRP) has gained prominence for its precise adaptation to mandibular contours and reported enhanced mechanical reliability compared to the manual-bent mandibular reconstruction plate (MBMRP). Despite this, a biomechanical comparison between MBMRP and PSMRP is essential for informed clinical decision-making and advancing the field of mandibular reconstruction. Thereupon, biomechanical behavior was compared between two reconstructed mandibles with the MBMRP made of commercial pure titanium plate and the PSMRP stemmed from a common ramus reconstruction case, respectively, in physiological simulation using finite element analyses. Results reveal that the PSMRP presents greater safety and better stability to the reconstructed mandible than the MBMRP in the physiological situation, indicated by over 45% lower peak equivalent von Mises stress and more than 35% less maximum displacement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Imaging & Pathology, Faculty of Medicine, KU Leuven, Kapucijnenvoer 7, Box 7003, Leuven 3000, Belgium;Oral and Maxillofacial Surgery, University Hospitals Leuven, Kapucijnenvoer 33, Leuven 3000, Belgium;Ch…
- Department of Materials Engineering, KU Leuven, Elooi Laboratory, Gent Campus, Gebroeders de Smetstraat 1, Gent 9000, Belgium;Collaborative Innovation Center of Steel Technology, University of Science and Technology Bei…
- KU Leuven.
- Chongqing Institute of Green and Intelligent Technology.
- Department of Imaging & Pathology, Faculty of Medicine, KU Leuven, Kapucijnenvoer 7, Box 7003, Leuven 3000, Belgium;Oral and Maxillofacial Surgery, University Hospitals Leuven, Kapucijnenvoer 33, Leuven 3000, Belgium.
- Oral and Maxillofacial Surgery, University Hospitals Leuven, Kapucijnenvoer 33, Leuven 3000, Belgium.
Zitieren
APA
Zhong, S., Zhang, Y., Shi, Q., Van Dessel, J., Geusens, J., Coppieters, S., Van Meerbeek, B., Politis, C., Yang, S., & Sun, Y. (2025). Biomechanics of Manual-Bent Versus Patient-Specific Mandibular Implants. Journal of Biomechanical Engineering. https://doi.org/10.1115/1.4069665
Vancouver
Zhong S, Zhang Y, Shi Q, Van Dessel J, Geusens J, Coppieters S, et al. Biomechanics of Manual-Bent Versus Patient-Specific Mandibular Implants. Journal of Biomechanical Engineering. 2025. doi:10.1115/1.4069665
BibTeX
@article{Zhong2025,
author = {Zhong, Shengping and Zhang, Yi and Shi, Qimin and Van Dessel, Jeroen and Geusens, Joris and Coppieters, Sam and Van Meerbeek, Bart and Politis, Constantinus and Yang, Shoufeng and Sun, Yi},
title = {Biomechanics of Manual-Bent Versus Patient-Specific Mandibular Implants},
journal = {Journal of Biomechanical Engineering},
year = {2025},
doi = {10.1115/1.4069665}
}