Fachartikel in der Zeitschrift Journal of Biomechanical Engineering, 2025. Verfasst von Brendan M Curran und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, University of Denver, 2155 East Wesley Avenue, Denver, CO 80208. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Shoulder, Mechanical Phenomena, Shoulder Joint, Models, Biological. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAMS NIH HHS, National Institute of Arthritis and Musculoskeletal and Skin Diseases.
Themen: Mobilisation
Abstract (Originalsprache)
Several complications are associated with total shoulder arthroplasty (TSA), including glenoid loosening and joint instability. These complications are linked to imbalances in glenohumeral muscle forces, which cause eccentric loading that occurs on the outer edge of the glenoid component, rather than in the center of the glenoid, and can be damaging to the implant and destabilizing the joint. The purpose of this study was to develop musculoskeletal models of TSA patients using their skeletal geometry acquired from CT and their kinematics acquired with high-speed stereo radiography (HSSR) to predict glenohumeral joint loading and muscle forces for abduction and flexion and describe how these loads differ from their native contralateral side. CT data was collected for both shoulders of 6 TSA patients (three anatomic and three reverse). Models were developed using CT reconstructions of each patient’s scapula and humerus, and placed implant geometry with CT-derived alignment. Mesh morphing was used to create personalized muscle attachment sites on the scapula. Simulation of abduction and flexion to quantify muscle and joint forces was driven with limb-specific HSSR kinematics. At 90 deg abduction, reverse TSA increased superior shear force by 11.2% bodyweight (BW) and decreased medial compressive force by 27.5%BW compared to each patient’s contralateral shoulder. Lateral deltoid elevation moment arms and stability ratios increased 9.61 mm and 0.58, respectively. At 90 deg abduction, anatomic TSA patients increased humeral internal rotation by 52.7 deg and decreased rotator cuff force by 9.4%BW. Patient-specific musculoskeletal models quantified differences in glenohumeral joint mechanics. This study revealed muscle and kinematic mechanisms that could lead to edge loading and glenoid loosening in TSA and can inform personalized surgical planning for implant alignment with the goal of improving patient outcomes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Mechanical Engineering, University of Denver, 2155 East Wesley Avenue, Denver, CO 80208.
- University of Denver.
- Mayo Clinic, Assistive and Restorative Technology Laboratory, 200 1st Street SW Rochester, MN 55905.
- Mayo Clinic.
ORCID-Profile: Thor E Andreassen
Förderung
- NIAMS NIH HHS
- National Institute of Arthritis and Musculoskeletal and Skin Diseases
Zitieren
APA
Curran, B. M., Myers, C. A., Laz, P. J., Andreassen, T. E., & Shelburne, K. B. (2025). Contributions of Muscle Forces to Stability and Mobility in Radiography-Driven Models of Total Shoulder Arthroplasty. Journal of Biomechanical Engineering. https://doi.org/10.1115/1.4069022
Vancouver
Curran BM, Myers CA, Laz PJ, Andreassen TE, Shelburne KB. Contributions of Muscle Forces to Stability and Mobility in Radiography-Driven Models of Total Shoulder Arthroplasty. Journal of Biomechanical Engineering. 2025. doi:10.1115/1.4069022
BibTeX
@article{Curran2025,
author = {Curran, Brendan M and Myers, Casey A and Laz, Peter J and Andreassen, Thor E and Shelburne, Kevin B},
title = {Contributions of Muscle Forces to Stability and Mobility in Radiography-Driven Models of Total Shoulder Arthroplasty},
journal = {Journal of Biomechanical Engineering},
year = {2025},
doi = {10.1115/1.4069022}
}
EUVIMED