Biomechanical Evaluation of Tracheal Needle Puncture Forces: Comparative Analysis of Annular Ligaments and Tracheal Cartilage
Journal of Biomechanical Engineering · 2024
DOI: 10.1115/1.4063821 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Biomechanical Engineering, 2024. Verfasst von Cheick Sissoko und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology-Head & Neck Surgery, University of Cincinnati College of Medicine, 231 Albert Sabin Way, ML 0528, Cincinnati, OH 45267. Hauptthemen laut MEDLINE: Trachea, Tracheostomy. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 34 Literaturangaben. Gefördert von: Air Force Office of Scientific Research, Chief Scientist Office.
Abstract (Originalsprache)
Percutaneous tracheotomies (PCT) are commonly performed minimally invasive procedures involving the creation of an airway opening through an incision or puncture of the tracheal wall. While the medical intervention is crucial for critical care and the management of acute respiratory failure, tracheostomy complications can lead to severe clinical symptoms due to the alterations of the airways biomechanical properties/structures. The causes and mechanisms underlaying the development of these post-tracheotomy complications remain largely unknown. In this study, we aimed to investigate the needle puncture process and its biomechanical characteristics by using a well establish porcine ex vivo trachea to simulate the forces involved in accessing airways during PCT at varying angular approaches. Given that many procedures involve inserting a needle into the trachea without direct visualization of the tracheal wall, concerns have been raised over the needle punctures through the cartilaginous rings as compared to the space between them may result in fractured cartilage and post-tracheostomy airway complications. We report a difference in puncture force between piercing the cartilage and the annular ligaments and observe that the angle of puncture does not significantly alter the puncture forces. The data collected in this study can guide the design of relevant biomechanical feedback system during airway access procedures and ultimately help refine and optimize PCT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology-Head & Neck Surgery, University of Cincinnati College of Medicine, 231 Albert Sabin Way, ML 0528, Cincinnati, OH 45267.
- University of Cincinnati College of Medicine, 231 Albert Sabin Way, ML 0528, Cincinnati, OH 45267.
Förderung
- Air Force Office of Scientific Research
- Chief Scientist Office
Zitieren
APA
Sissoko, C., Walker, V., & Dion, G. R. (2024). Biomechanical Evaluation of Tracheal Needle Puncture Forces: Comparative Analysis of Annular Ligaments and Tracheal Cartilage. Journal of Biomechanical Engineering. https://doi.org/10.1115/1.4063821
Vancouver
Sissoko C, Walker V, Dion GR. Biomechanical Evaluation of Tracheal Needle Puncture Forces: Comparative Analysis of Annular Ligaments and Tracheal Cartilage. Journal of Biomechanical Engineering. 2024. doi:10.1115/1.4063821
BibTeX
@article{Sissoko2024,
author = {Sissoko, Cheick and Walker, Victoria and Dion, Gregory R},
title = {Biomechanical Evaluation of Tracheal Needle Puncture Forces: Comparative Analysis of Annular Ligaments and Tracheal Cartilage},
journal = {Journal of Biomechanical Engineering},
year = {2024},
doi = {10.1115/1.4063821}
}