A Novel Approach to Open Reduction and Internal Fixation of Distal Radius Fractures Utilizing a Multi-Degree-of-Freedom Traction and Stabilization Device
Journal of Medical Devices · 01.06.2022 · 16 (2)
DOI: 10.1115/1.4052901 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Devices, online erschienen am 03.02.2022. Verfasst von Jacob F Oeding und 5 weiteren Autor:innen. Beteiligte Einrichtung: New York University Grossman School of Medicine, New York, NY 10016; Department of Mechanical Engineering, Massachusetts Institute of Technology, Cam…. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 4 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Open reduction and internal fixation (ORIF) is a surgical procedure performed with the objectives of restoring normal alignment and providing stability to broken bone fragments after a fracture. This procedure is increasingly used to treat fractures of the distal end of the radius. Reduction is achieved by the surgeon pulling and manipulating the hand while looking at real-time X-rays, and frequently requires large forces to distract impacted fragments from the proximal bone. This study presents the design and preliminary testing of a multi-degree-of-freedom (DOF) device capable of performing both distraction and reduction of fractured bone fragments using a traction splint mechanism with locking ball joints. A prototype was manufactured, and tests were conducted by a practicing hand surgeon. Both qualitative and quantitative tests using a phantom arm were performed. Quantitative force testing found an 80% reduction in the maximum force required to create needed traction, while qualitative tests with a hand surgeon found the device's ability to reduce and stabilize bone fragments while the hardware is secured to be more intuitive and less obstructive than existing techniques.
Autor:innen und Institutionen
- New York University Grossman School of Medicine, New York, NY 10016; Department of Mechanical Engineering, Massachusetts Institute of Technology, Cambridge, MA 02139.
- Department of Mechanical Engineering, Massachusetts Institute of Technology, Cambridge, MA 02139.
- Beth Israel Lahey Health Mount Auburn Hospital Orthopedics, Cambridge, MA 02138.
Zitieren
APA
Oeding, J. F., Bockman, S., Chiu, H., Hua, C., Connor, J., & Slocum, A. (2022). A Novel Approach to Open Reduction and Internal Fixation of Distal Radius Fractures Utilizing a Multi-Degree-of-Freedom Traction and Stabilization Device. Journal of Medical Devices, 16(2). https://doi.org/10.1115/1.4052901
Vancouver
Oeding JF, Bockman S, Chiu H, Hua C, Connor J, Slocum A. A Novel Approach to Open Reduction and Internal Fixation of Distal Radius Fractures Utilizing a Multi-Degree-of-Freedom Traction and Stabilization Device. Journal of Medical Devices. 2022;16(2). doi:10.1115/1.4052901
BibTeX
@article{Oeding2022,
author = {Oeding, Jacob F and Bockman, Sam and Chiu, Harriet and Hua, Cheyenne and Connor, James and Slocum, Alexander},
title = {A Novel Approach to Open Reduction and Internal Fixation of Distal Radius Fractures Utilizing a Multi-Degree-of-Freedom Traction and Stabilization Device},
journal = {Journal of Medical Devices},
year = {2022},
volume = {16},
number = {2},
doi = {10.1115/1.4052901}
}
EUVIMED