Accuracy of next‐generation wireless skin temperature sensors during exercise–heat stress

Aaron J E Bach iD, Norman R Morris iD, Surendran Sabapathy iD, David N Borg iD, Isabella Ennever iD, Pramod Sharma iD, Fergus K O'Connor iD

Experimental Physiology · 2026

DOI: 10.1113/ep093856 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Experimental Physiology, 2026. Verfasst von Aaron J E Bach und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Health Sciences and Social Work, Griffith University, Gold Coast, Queensland, Australia. Hauptthemen laut MEDLINE: Skin Temperature, Exercise, Wireless Technology, Heat-Shock Response. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Skin temperature is fundamental in characterising human thermoregulatory responses. Wired probes, although accurate, restrict movement and are impractical outside laboratory settings. The iButton (DS1922L; Maxim Integrated, USA) is a widely used wireless alternative but does not meet the precision recommended by ISO 9886 and permits only retrospective data retrieval. A recently developed wireless sensor, the eTemp Performance (BodyCAP, France), claims to address these limitations but has not been validated independently. Here, we compared the eTemp and iButton against a wired thermocouple reference (SST‐1; Physitemp, USA) during rest, exercise and recovery in the heat [mean (SD): 35.0 (0.5)°C, 40.6 (1.5)% relative humidity]. Twenty‐six adults (10 women) completed seated rest (15 min), cycling at fixed heat production (30 min) and passive recovery (15 min). Mean skin temperature was calculated using the Ramanathan four‐site formula. Across all periods, the eTemp showed a small negative bias (mean difference, −0.04°C; 95% limits of agreement, −0.33°C to 0.25°C; concordance correlation coefficient, 0.986), whereas the iButton showed a small positive bias (+0.11°C; 95% limits of agreement, −0.23°C to 0.46°C; concordance correlation coefficient, 0.974). The eTemp met all a priori acceptability thresholds for mean skin temperature. The iButton met the acceptability threshold for mean bias in all periods and was statistically equivalent to the reference; however, the 95% confidence interval on the upper limit of agreement marginally exceeded 0.5°C during exercise, recovery and overall. These findings provide the first independent validation of the eTemp Performance sensor and support the use of both wireless devices for mean skin temperature measurement during exercise–heat stress, notwithstanding the known limitations in accuracy of the iButton against the ISO 9886 standard.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Skin TemperatureExerciseWireless TechnologyHeat-Shock ResponseHumansFemaleMaleAdultBody Temperature RegulationHot TemperatureYoung Adult

Stichwörter der Autor:innen

ambulatory monitoringbody temperature regulationergometryhot temperaturereproducibility of resultswearable electronic device

Autor:innen und Institutionen

  • School of Health Sciences and Social Work, Griffith University, Gold Coast, Queensland, Australia.
  • Australia Centre for Precision Health and Technology, Griffith University, Gold Coast, Queensland, Australia.
  • Griffith Institute for Human and Environmental Resilience, Griffith University, Nathan, Queensland, Australia.
  • Allied Health Research Collaborative, The Prince Charles Hospital, Metro North Hospital and Health Service, Chermside, Queensland, Australia.
  • School of Health, University of the Sunshine Coast, Sippy Downs, Queensland, Australia.
  • School of Pharmacy and Medicine, Griffith University, Gold Coast, Queensland, Australia.

ORCID-Profile: Aaron J E Bach, Norman R Morris, Surendran Sabapathy, David N Borg, Isabella Ennever, Pramod Sharma, Fergus K O'Connor

Zitieren

APA

Bach, A. J. E., Morris, N. R., Sabapathy, S., Borg, D. N., Ennever, I., Sharma, P., & O'Connor, F. K. (2026). Accuracy of next‐generation wireless skin temperature sensors during exercise–heat stress. Experimental Physiology. https://doi.org/10.1113/ep093856

Vancouver

Bach AJE, Morris NR, Sabapathy S, Borg DN, Ennever I, Sharma P, et al. Accuracy of next‐generation wireless skin temperature sensors during exercise–heat stress. Experimental Physiology. 2026. doi:10.1113/ep093856

BibTeX
@article{Bach2026,
  author = {Bach, Aaron J E and Morris, Norman R and Sabapathy, Surendran and Borg, David N and Ennever, Isabella and Sharma, Pramod and O'Connor, Fergus K},
  title = {Accuracy of next‐generation wireless skin temperature sensors during exercise–heat stress},
  journal = {Experimental Physiology},
  year = {2026},
  doi = {10.1113/ep093856}
}
Kostenlos starten →