Ex vivo biomechanical testing of a three‐dimensional printed titanium plate and spacer construct and 4.5 mm locking compression plate for ventral cervical fusion of C4 – C5 in the horse
Veterinary Surgery · 2025
DOI: 10.1111/vsu.70014 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Surgery, 2025. Verfasst von Steven Zedler und 5 weiteren Autor:innen. Beteiligte Einrichtung: University of Queensland, School of Veterinary Science, Gatton, Queensland, Australia. Hauptthemen laut MEDLINE: Bone Plates, Cervical Vertebrae, Spinal Fusion. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Objective
To compare the biomechanical properties of a three‐dimensional (3D) printed titanium plate and spacer (3DM) and a 4.5 mm broad locking compression plate (LCP) for intervertebral fusion of C4–C5 in the horse. Study design Ex vivo experimental study. Animals Twenty‐four equine cadaveric cervical spines.
Methods
Twenty‐four cadaver equine cervical spines were randomly assigned to ventral intervertebral fusion with either 3DM ( n = 12) or LCP ( n = 12) at C4–C5. Four‐point bending tests were performed in flexion (3DM n = 6, LCP n = 6) and extension (3DM n = 6, LCP n = 6) in a single cycle to failure. Yield, failure moment, and stiffness in flexion and extension were compared between the 3DM and LCP groups using a mixed effects linear regression model. Failure mode was compared using penalized logistic regression. The significance level was p ≤ .05.
Results
Yield, failure moment, and stiffness were similar between LCP and 3DM groups, with no differences in either flexion or extension. In extension, displaced fractures ( p = .03) and screw pullout ( p = .09) were more common in the LCP group.
Conclusion
The 3DM group had similar yield, failure moment, and stiffness to the LCP group. The LCP constructs were more likely to fail by screw pullout or displaced vertebral fracture in extension. Sample size was a major limitation, so smaller differences in the mechanical variables and failure mode would not have been detected. Clinical significance In a cadaver model, the 3DM construct was less likely to fail by displaced fracture when loaded in extension. This warrants in vivo investigation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- University of Queensland, School of Veterinary Science, Gatton, Queensland, Australia.
- Department of Clinical Studies-New Bolton Center, School of Veterinary Medicine, University of Pennsylvania, Kennett Square, Pennsylvania, USA.
- Clinique Equine de Grosbois, Boissy Saint Léger, France.
Zitieren
APA
Zedler, S., Jukic, C., van Eps, A., Stefanovski, D., Genton, M., & Rossignol, F. (2025). Ex vivo biomechanical testing of a three‐dimensional printed titanium plate and spacer construct and 4.5 mm locking compression plate for ventral cervical fusion of C4 – C5 in the horse. Veterinary Surgery. https://doi.org/10.1111/vsu.70014
Vancouver
Zedler S, Jukic C, van Eps A, Stefanovski D, Genton M, Rossignol F. Ex vivo biomechanical testing of a three‐dimensional printed titanium plate and spacer construct and 4.5 mm locking compression plate for ventral cervical fusion of C4 – C5 in the horse. Veterinary Surgery. 2025. doi:10.1111/vsu.70014
BibTeX
@article{Zedler2025,
author = {Zedler, Steven and Jukic, Chantelle and van Eps, Andrew and Stefanovski, Darko and Genton, Martin and Rossignol, Fabrice},
title = {Ex vivo biomechanical testing of a three‐dimensional printed titanium plate and spacer construct and 4.5 mm locking compression plate for ventral cervical fusion of C4 – C5 in the horse},
journal = {Veterinary Surgery},
year = {2025},
doi = {10.1111/vsu.70014}
}
EUVIMED