Effect of suture anchor type, eyelet configuration, and loading condition on suture failure: An in vitro study

Obinna P Fidelis, Pierre-Yves Mulon, David E Anderson, Dustin L Crouch

Veterinary Surgery · 2025

DOI: 10.1111/vsu.14231 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Surgery, 2025. Verfasst von Obinna P Fidelis und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical, Aerospace and Biomedical Engineering, University of Tennessee, Knoxville, Tennessee, USA. Hauptthemen laut MEDLINE: Suture Anchors, Sutures, Suture Techniques. Die Arbeit nennt 46 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAMS NIH HHS, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To determine the effect of suture anchor type, eyelet configuration (raised or embedded), and loading condition (without or with cyclic loading) on failure load and failure mode of sutures in anchors with embedded and raised eyelets. Study design Controlled laboratory study. Sample population A total of 4 unique clinical anchor+suture combinations ( n = 40 total); eyebolt screw reference samples ( n = 30 total).

Methods

A total of 10 samples of each anchor+suture combination were either (a) loaded to failure ( n = 5) or (b) loaded from 0 N to 60 N for 1000 cycles then loaded to failure ( n = 5). As a “best‐case” reference, an eyebolt screw with a smooth, round cross‐section was tested under the same two loading conditions and the same sutures ( n = 10 per suture) used with the clinical anchors. The percent difference in the maximum load at failure () between each anchor and suture‐matched eyebolt screw was used to quantify the effect of each anchor on suture strength.

Results

There was no effect of loading condition (i.e., without or with cyclic loading) on F max . The Anika anchor had a significantly lower than the Arthrex anchor ( p = .015), IMEX anchor ( p = .004) and Jorvet anchor ( p < .001). A total of 14 out of 20 samples of anchors with raised eyelets failed at the suture mid‐section, whereas all 20 samples of anchors with embedded eyelets failed at the suture knot.

Conclusion

Anchors with raised eyelets had more failures at the suture mid‐section indicative of wear and/or cutting of the suture against the eyelet. Clinical significance Eyelet geometry and surface finish of metallic screw‐type anchors have effects on suture strength.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Suture AnchorsSuturesSuture TechniquesAnimalsMaterials TestingTensile StrengthEquipment Failure

Autor:innen und Institutionen

  • Department of Mechanical, Aerospace and Biomedical Engineering, University of Tennessee, Knoxville, Tennessee, USA.
  • Department of Large Animal Clinical Sciences, College of Veterinary Medicine, University of Tennessee, Knoxville, Tennessee, USA.

Förderung

  • NIAMS NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Fidelis, O. P., Mulon, P. Y., Anderson, D. E., & Crouch, D. L. (2025). Effect of suture anchor type, eyelet configuration, and loading condition on suture failure: An in vitro study. Veterinary Surgery. https://doi.org/10.1111/vsu.14231

Vancouver

Fidelis OP, Mulon PY, Anderson DE, Crouch DL. Effect of suture anchor type, eyelet configuration, and loading condition on suture failure: An in vitro study. Veterinary Surgery. 2025. doi:10.1111/vsu.14231

BibTeX
@article{Fidelis2025,
  author = {Fidelis, Obinna P and Mulon, Pierre-Yves and Anderson, David E and Crouch, Dustin L},
  title = {Effect of suture anchor type, eyelet configuration, and loading condition on suture failure: An in vitro study},
  journal = {Veterinary Surgery},
  year = {2025},
  doi = {10.1111/vsu.14231}
}
Kostenlos starten →